The Settings/Sound Setup menu allows you to
set the limit values of the pulse rate per minute at
which acoustic (acoustic signal) and visual
alarms (blinking of the measurement) are
triggered. In the pulse rate per minute range of 5
- 250, a lower and an upper limit can be set.
However, as the device can only measure a pulse
rate of more than 30 beats per minute, the
minimum limit should be set to at least 30 beats
per minute. To adjust the limits, switch on the
device and hold down the M configuration
button until the Settings/Sounds Setup mode is
displayed and then release the button . The
star-shaped cursor is now displayed next to
Sounds Setup. Briefly press the M
configuration button three times to
8.5 Pulse rate alarm settings
Release the button when the desired value is
reached and move the star-shaped cursor down to
the SpO2 Alm Lo entry by pressing the button
again. Adjust the lower alarm limit next to the
star-shaped cursor in the same way as the upper
alarm limit. Make sure that the upper limit is
higher than the lower limit. To save the selected
settings, move the star-shaped cursor to the Exit
entry by pressing the configuration button and
hold the button down until the measurements
are displayed again. The settings are now saved.