1
4 white digit Display
Shows value according configuration.
Display will flash wheen relay output is activated.
Shows steps and data during configuration.
2
Minus sign LED
It iluminates for negative readings.
It iluminates for negative values.
3
Keyboard
—
—
4
UP key
Increases the value of the setpoint generator if it is
previously enabled in 'diMM' mode or increases the active
digit if it is enabled in 'ProG' mode.
Shows Setpoints value.
Increases value of active digit.
5
SHIFT / DOWN key
Displays the maximum and minimum values when the
setpoint generator is not enabled.
Update the maximum and / or minimum value stored to the
current display value if you press more than 3s.
Decreases the value of the setpoint generator if it is enabled
in 'diMM' mode.
Shifts flashing active digit to the next right digit.
Shows sequentially menu options.
6
DATA/ENTER key
Switch to PRO mode.
Followed by pressing 'DOWN', enters to configure the value
of the setpoint generator for the 'ProG' mode.
Followed by pressing 'DOWN' for 5s, it allows to disable /
enable the adjustment modes 'diMM' and 'ProG' of the
setpoint generator.
Validates selected data and parameters.
Moves one step forward in configuration menu.
Changes to RUN mode.
7
Free space for units
label
—
—
Installing and connecting recommendations
WARNING: If this instrument is not installed and used in accordance with this instructions,
the protection provided by it against hazards may be impaired.
To meet the requirements of EN 61010
-
1 standard, where the unit is permanently connected to main
supply, its is obligatory to install a circuit breaking device easy reachable to the operator and clearly
marked as the disconnecting device.
To guarantee electromagnetic compatibility, the following guidelines should be kept in mind:
•
Power supply wires should be separatedly routed from signal wires and
never runned
in the same
conduit.
•
Use shielded cable for signal wiring.
•
Cables section should be
0.25 mm²
.
Before connecting signal wires, signal type and input range should be verified to be within the right
limits.
Do not connect simultaneously more than one input signal to the meter
.
This instrument complies with the EU directives of EMC i LVD.
Refer to the instructions in this manual to preserve safety protections.
DOWN
There are two main function modes:
RUN
and
PRO
.
PRO
mode is when
configuration menu is entered to
programm the indicator, whereas
RUN
is the normal mode in which
display shows the reading according
to configuration and input signal
value.
The table below summarizes display
parts description and LEDs and
keyboard function.