background image

Switch on the system with the mains switch on the back of the system. The 

start menu will be displayed. The green LED turns on as soon as a process is 

running. 

Operation 

Parameter Setting 

Start menu 

 

 

This menu is displayed after the power-on, 

after abortion of after the normal completion 

of a process. 

The last set values are retained even after the 

switch-off of the system. 

Setup menu

 

 

 

Push the 

up arrow key for 5 seconds

, then 

the setup menu is displayed. With 

+

 

and 

 

the values can be selected for setting. 

The selected box flashes, with 

ok

 it will be 

permanently set for changing. With the keys 

+

 and 

-

,

 the input or the values can be 

changed. 

If an input is confirmed with 

ok

, the box is 

highlighted flashing again. 

With confirming the button 

Start menu

, the 

start menu is displayed again. 

 

The following values can be adapted: 

display (-colour) 

contrast 

brightness 

language 

rotation direction 

confirm process end: 

Yes: at the process end, a signal is audible 

Summary of Contents for HLC MHL 23

Page 1: ...1 Thermo Shaker and BlockThermostats with Smart Control MKR 13 MKR 23 MHR 13 MHR 23 MHL 23 TK 23 Operating Manual...

Page 2: ...4 Temperature accuracy 5 First Steps 5 Setup 5 Operation 6 Parameter Setting 6 Working with the Programming Function 10 Assembly of Accessories 13 Anti Condensation Plate BA 24 96 800012900 800013000...

Page 3: ...s must be avoided The ventilation of the device must not be blocked To ensure the proper ventilation at all times please consider a mini mum distance of 10 centimeter on each side of the device Damage...

Page 4: ...extrac tor hood If the ThermoMixers are operated with a high shaking frequency the vibrations of the system might transmit to the underground Please do consider this when placing the system Contact wi...

Page 5: ...red When setting a process time please note that the sample temperature is always reached a little later than the displayed block temperature First Steps Setup Plug the mains cable in the IEC socket o...

Page 6: ...stem Setup menu Push the up arrow key for 5 seconds then the setup menu is displayed With and the values can be selected for setting The selected box flashes with ok it will be permanently set for cha...

Page 7: ...hange can also be made in the same way during a running process After confirming the input with ok the new set value is effective immediately For the hundreds the following values can be selected 0 fo...

Page 8: ...endless process select 00 00 100 Short Mix vortexing Navigate with and to the button 100 Short Mix and confirm with ok The system shakes with maximum mixing speed regardless of a process is started or...

Page 9: ...emaining time is displayed in the menu The time only starts to run if the temperature set value is reached This does also apply for subsequent tempera ture changes A set value change can also be made...

Page 10: ...layed after the expiration of the time The device turns off automatically Working with the Programming Function Program menu Navigate with and to the button Pro grams and confirm with ok A total of 30...

Page 11: ...t ting is only required if the sample is to be tempered slower than technically possi ble If the sample is to be tempered as fast as possible set the value to 0 Rotational speed Shaking frequency with...

Page 12: ...ng program The set time is the time of the running step the remaining time is the time of the complete process In the lower right part of the display program no cycle and step are indicated Terminatio...

Page 13: ...hold of the bolt of the system and with that tightly attaches block and anti condensation to the system Data Transfer A USB 1 1 interface for the communication with a PC is provided as a stand ard Con...

Page 14: ...hh mm tmp rpm prg cycle step steptime hh mm rem risetime_m mix_s pause_m 00 00 26 9 0000 01 01 01 00 00 00 9 0 2 00 01 33 7 0220 01 01 01 00 00 00 9 0 2 00 02 36 7 0000 01 01 01 00 00 00 9 0 2 00 03 3...

Page 15: ...the next valid value r 020 Changes the set rotational speed to a new value The rotational speed must always be entered with three digits r 120 sets 1 200 1 min r 045 sets 450 1 min r is the identific...

Page 16: ...correct system type name at power on If not please contact the HLC by DITABIS service department There are excessive temperature fluctuations Check the seat of the changeable block by pulling it up If...

Page 17: ...slit Do not perform spray disinfection Do only connect the system with the power supply if it is completely dry The repair service may only be performed by staff authorized and trained by the manufac...

Page 18: ...off all outer parts of the system with the disinfectant and a lint free cloth In Case of Service If a technical problem arises please contact your local dealer or the DITABIS service department The D...

Page 19: ...ity 1 changeable block Weight 9 kg Electr heating cooling power 130 W Electr supply 230 V 50 Hz Technical Data MKR 23 Temperature working range Room temp 11 C to 70 C Temperature adjustable range 0 C...

Page 20: ...y 1 changeable block Weight 7 kg Electr heating cooling power 200 W Electr supply 230 V 50 Hz Technical Data MHR 23 Temperature working range Room temp 3 C to 130 C Temperature adjustable range 0 C to...

Page 21: ...ity 2 changeable blocks Weight 7 kg Electr heating cooling power 350 W Electr supply 230 V 50 Hz Technical Data TK 23 Temperature working range Room temp 16 C to 90 C Temperature adjustable range 10 C...

Page 22: ...deep flat bottom 800011900 BP 28 for 8 x D 28 mm 40 mm deep flat bottom For centri 800012200 BZ 15 Falcon tubes 14 x 15 ml with isolation lid fuge tubes 800012300 BZ 50 for Falcon tubes 6 x 50 ml wit...

Page 23: ...T 01 Tempering tub for 1 block transparent 800013900 BT 02 Tempering tub for 2 blocks transparent 800012900 BA 24 Anti condensation plate for 1 block micro test tubes 800013000 BA 96 Anti condensation...

Page 24: ...24 Explanations Caution Risk of electric shock Caution Caution Hot surface Important information Earth conductor Fuse i...

Page 25: ...25 800 68 13 02_Manual_Smart_Control docx...

Page 26: ...26...

Reviews: