Operating Instructions UVD 170S/340S
9
4.3 Installation and Operation within CHROMELEON
TM
4.3.1 Communication Port
The
UVD 170S/340S
is controlled via a special PC-interface board (
→
4.3.2 Required CHROMELEON
TM
Modules
To be able to control the detector
UVD 170S/340S
via CHROMELEON
TM
, the corresponding
CHROMELEON
TM
Control Option is required. For questions, please contact your Dionex
sales representative.
4.3.3 Installation with the CHROMELEON
TM
Installation Tool
Proceed as follows to install the detector
UVD170S/340S
in the CHROMELEON
TM
Installation Tool:
•
Start the CHROMELEON
TM
-Server and then the CHROMELEON
TM
Installation Tool
(= Server Configuration).
•
Select the timebase to which you want to assign the detector
UVD170S/340S
.
•
Via the Edit menu or the menu of the right mouse button, choose "
Add Device...
".
•
Choose "
Dionex UVD170S
" or "
Dionex UVD340S
" from the displayed list. Confirm by
pressing "
OK
".
The now displayed tab dialog boxes show the current configuration of your detector.
Generally, these settings must not be changed. However, please check whether the settings
correspond to your current installation environment. If necessary, change settings.
Tab Dialog Box "General"
Option Description
Device Name
Instrument name under which the detector is listed in the
installation environment (default: UV)
Port Address
I/O-Base address of the PC interface board (default: 318H)
Demo Mode
Deactivate the demo mode!