MCW-MCWR2-MCWK PROFESSIONAL MCW-MCWK_03_12.10_EN_U
37
3.2.2 Regular Maintenance
Maintenance should be carried out only by persons who have acquired the necessary technical expertise and are
specialized and trained for this purpose.
Every 3 months or, in any case, after 12.500
weighs
-
Check all dimensions of the parts which make up the instrument;
-
Check the wear on the handle or the eyelet, by checking if there
are any plastic deformations, mechanical damages(irregular),
cracks, corrosion, damage to threaded portions and the twists;
-
Check the tightness of the splice plate on the hook, and the
presence of defects, and ensure its proper functioning;
-
Make sure that the split pin and the shackle nuts are firm;
-
If other metrological and mechanical irregularities are detected,
have the electronic crane scale repaired by qualified personnel
(Dini Argeo assistance service).
Do not for any reason carry out the repair by yourself.
In case of non-compliance turn immediately off the electronic crane
scale.
All repair operations and the parts used are classified and stored in
the maintenance register.
Every 12 months or, in any case, after 50.000
weighs
The extraordinary maintenance of the product should be made by
specialized personnel (Dini Argeo assistance service).
It is of utmost importance that all the maintenance and repair operations, and the used pieces are recorded
and filed in the appropriate produce maintenance registry.