Advanced Force Gauge User’s Manual
20
Example 5
Settings: -
- BOTH LED and audio
alarms are active
- Alarm triggers on OUT
BAND
- Alarm is set to FAIL
- % DROP is 10% of
fullscale (e.g. AFG
100N must register
drop of 10N)
Main display is set to
1st peak tension
screen
PLC
(Programmable Limit
Controller)
PLC sub-menu 1
The AFG has a load output signal which may be used
for PLC applications. For PLC applications, this
function requires an external cable with a built-in
solid-state relay – see
Specifications
for details of the
signal.
To configure the signal output from the AFG, press
and hold the
MENU
key until page 1 of the main menu
appears. Press
DOWN
to move the arrow cursor to
PLC
and press the
ENTER
key. The cursor arrow now
points to
PLC OFF
.
The display will show:
PLC OFF
Indicates PLC function status.
RESET
When the load limit is reached, the
output signal triggers the relay and
the RESET key must be pressed to
clear the line before starting the next
test.
CONTINUOUS The relay will be activated every time
the load limit is reached and the
output signal will remain on.
PULSE
The relay will be activated momen-
tarily when the load limit is reached.
Select the desired function and press the
ENTER
key.
Summary of Contents for Quantrol AFG 10
Page 1: ...Advanced Force Gauge AFG AFTI User s Manual ...
Page 4: ...Advanced Force Gauge User s Manual 4 ...
Page 34: ...Advanced Force Gauge User s Manual 34 Stand ...
Page 35: ...Advanced Force Gauge User s Manual 35 Freeze Drop Average Time Rate ...
Page 36: ...Advanced Force Gauge User s Manual 36 Footswitch1 Footswitch2 Comms Constant ...
Page 37: ...Advanced Force Gauge User s Manual 37 Information Calibration Contrast ...
Page 38: ...Advanced Force Gauge User s Manual 38 ...