Advanced Force Gauge User’s Manual
28
The transmission (or Baud) rate can now be set.
Use the
UP
or
DOWN
key to position the arrow cursor
at the relevant speed (2400, 9600, 57600 or 115200).
Press ENTER to select.
You will now return to page 2 of the main menu.
To set STORE MEM press
ENTER
from Comms sub-
menu 1. This will cause a memory counter to appear
in the main display. You will now return to page 2 of
the mainmenu.
To set SEND MEM press
ENTER
from Comms sub-
menu 1. This will cause a TX symbol to flash in the
main display as the memory data is now transmitted
to a peripheral device. The data is transmitted at the
settings defined by PORT. After transmission of data
you will now return to page 2 of the main-menu.
To set CLEAR MEM press
ENTER
from Comms sub-
menu 1. This now erases all the data stored in the
memory. The memory counter is now reset to zero.
After clearing the memory you will now return to page
2 of the main-menu.
Displays calibration information.
T - Tension span
C - Compression span
Z - Current zero
G - Gravitational constant
This is for information only and may be required for
diagnostic purposes by your distributor.
Calibration menu - displays a password menu and
allows the user to check for load cell offset if overload
is suspected.
Place the gauge (or your external SMART sensor)
horizontally on a flat level surface. By using UP or
DOWN keys select the CAL option and the display will
show 0000. Press
ENTER
four times (to enter 0000
as the password), a screen showing initial (at calibra-
tion) load cell offsets and current offset in % is
displayed.
The upper value shows the % offset when the gauge
was last successfully calibrated. The lower value
shows the % offset currently present on the gauge.
Port sub-menu 2
CALIBRATION
INFORMATION
Summary of Contents for Quantrol AFG 10
Page 1: ...Advanced Force Gauge AFG AFTI User s Manual ...
Page 4: ...Advanced Force Gauge User s Manual 4 ...
Page 34: ...Advanced Force Gauge User s Manual 34 Stand ...
Page 35: ...Advanced Force Gauge User s Manual 35 Freeze Drop Average Time Rate ...
Page 36: ...Advanced Force Gauge User s Manual 36 Footswitch1 Footswitch2 Comms Constant ...
Page 37: ...Advanced Force Gauge User s Manual 37 Information Calibration Contrast ...
Page 38: ...Advanced Force Gauge User s Manual 38 ...