To set-up the ET6 for your specific application, you 

might need to change some of the Dip Switches. 

The Dip Switches are located underneath the chart. 

Use a pen or small screw driver to flip the switches. 

Remember to install correct chart to match 

corresponding Dip Switch setting.

Channel 1

 

  Volts   

 

#1 Off

      mA 

 

 

#1 On

Channel 2 (Optional)

  Volts   

 

#2 Off

      mA 

 

 

#2 On

Time

  1 day   

 

#3 On

      7 day   

 

#3 Off

Vdc/mA

  0-5 Vdc or 0-20 mA  

#4 Off

  1-5 Vdc or 4-20 mA  

#4 On

  0 to 100 

 

#5 Off

   

 

 

#6 Off 

   

 

 

#7 Off

  0 to 10° (0 to 100° chart) 

#5 Off

   

 

 

#6 Off

   

 

 

#7 On

  0 to +14,  

 

#5 Off

   

 

 

#6 On

   

 

 

#7 Off

  0 to 200 

 

#5 Off

   

 

 

#6 On

   

 

 

#7 On

  0 to 300 

 

#5 On

   

 

 

#6 Off

   

 

 

#7 Off

  -10 to 10 

 

#5 On

   

 

 

#6 Off

   

 

 

#7 On

  -20 to 120 

 

#5 On

   

 

 

#6 On

   

 

 

#7 Off

  -50 to 50 

 

#5 On

   

 

 

#6 On

   

 

 

#7 On

Alarm Speaker

The alarm speaker may be activated or deactivated 

with the use of dip switch #8. 

  Deactivated  

 

#8 On

      Activated  

 

#8 Off

The Alarm speaker will sound during minimum, 

maximum or probe missing alarm conditions.

Note: 

Dip switch settings are only updated when 

the unit is turned on.

Figure 5

All switches are 

shown in off 

position.

Slide toggle to this side for on

Slide toggle to this side for off

6

Dip Switch Set-Up

O

N

1 2 3 4 5 6 7 8

Your instrument was carefully tested and calibrated 

before being shipped from the factory.  additional 

calibration is not required.  However, should cali-

bration be desired in the future, we recommend 

that it be at our lab, Call Customer Service at 

630-543-3747.  If you wish to calibrate yourself, 

follow these procedures:

1.     There are two phases of calibration.  During 

        the first phase, the recorder should be 

        calibrated using an accurate electronic signal 

        generator.  The second phase consists of 

        calibrating the recorder in conjunction with 

        the transmitter to be used.  When calibrating 

        with a transmitter, a compatible standard 

        should be used.  For example, use a highly 

        accurate temperature standard to calibrate the 

        ET6 using a temperature transmitter. 

 

2.     The calibration procedure is the same whether 

        calibrating the unit only, using an electrical 

        signal calibrator, or calibrating the unit in 

        conjunction with a transmitter.  When 

        calibrating with a transmitter, the calibration 

        must be performed using a separate standard 

        with which to adjust the unit and transmitter.

3.     To activate the calibration mode, turn the unit 

        off.  now press the ON/OFF key and the 

        ALARM key at the same time.  The unit is in 

        the “User Calibration” mode when the red 

        LED is on.  The display will show a “C” during 

        the warm up period and then will flash between 

        0 and F or 0 and C (depending on measurement 

        selected.)  Match the reading of the recorder 

        to the reading of the reference standard.
4.     To raise the unit of measurement, press the 

        CHART ROTATION (Up Arrow) key.  To lower 

        the unit of measurement press the HOME 

        (Down Arrow) key.  To ensure that you have 

        matched the standard, allow the pens to 

        stablilize for at least 30 seconds before exiting 

        calibration mode.  If your ET6 only has one 

        pen, skip to Step 6.
5.     Press the alarm button for 2 seconds to switch 

        from Pen 1 to Pen 2.  Repeat step 4 for Pen 2 

        temperature calibration.
6.     When calibration is complete simply press the 

        ON/OFF key to save the calibration settings.
Calibration is stored in memory even after you 

turn the unit off.  User calibration information will

not be lost if AC power fails.

Calibration

Summary of Contents for ET6

Page 1: ...ut Recorder Registradora Universal de Entrada Universaleingangs Recorder ET6 O P E R A T I O N M A N U A L M A N U A L D E F U N C I O N A M I E N T O M A N U A L D U T I L I S A T I O N DICKSON www d...

Page 2: ...2 F 0 to 50 C 0 to 90 RH non condensing Note Response time is slower when using battery power source Your ET6 recorder has been preset to operate using the most popular settings To change the range us...

Page 3: ...t Hub Instrument Anatomy 3 Figure 2 Panel Mount Holes Relay Contact Optional AC DC Adapter Key Hole Wall Mount Holes Panel Mount Holes Input Jack 0 5V 0 20MA Relay Contact Pen 1 Blue 2 Pen Unit only I...

Page 4: ...ft of the recorder Tighten screws in terminal block to hold in transmitter wires Repeat steps with second terminal block if using the ET6 2 pen model Pen Hysteresis a property that occurs when tension...

Page 5: ...til it is disabled or until the alarm condition no longer applies Alarm 2 pen ET6 Holding the ALARM key for 2 seconds will cause the unit to enter into the alarm set mode for Pen 1 Pressing and holdin...

Page 6: ...ted using an accurate electronic signal generator The second phase consists of calibrating the recorder in conjunction with the transmitter to be used When calibrating with a transmitter a compatible...

Page 7: ...the pen arm 3 Slide the new pen on the pen arm until the notch in the pen is secure to the pen arm 4 Remove the pen cap Insert on the Pen Cap Holder 5 Press the PEN HOME key to return the pen to reco...

Page 8: ...or products not manufactured by it or for any products modified by buyer or subject to misuse or neglect Symptom Cause Check Remedy Pen trace too Tip too sharp cartridge Sand tip moisten or replace pe...

Page 9: ...misor 3 Quite las cubiertas protectoras de estilete s y col quelas en el sujetador para cubiertas de estilete 4 Oprima la tecla ON OFF para encender el instrumento Diagrama identificando las teclas p...

Page 10: ...stilete Unidad de estilete Flecha de Cronometraje LED Teclado V rtice de diagrama Anatom a del Instrumento 10 Figura 2 Orificios para montaje en panel Contacto de rel opcional Adaptador CA CD Orificio...

Page 11: ...a ter minal para fijar los cables Repita el procedimiento con la segunda terminal si est usando el modelo ET6 con dos estiletes Estilete La hist resis una propiedad que ocurre cuando se aplica tensi n...

Page 12: ...aplique la condici n Alarma ET6 con dos estiletes Para encender la alarma del estilete 1 oprima la tecla de ALARMA durante dos segundos Para encender la alarma del estilete 2 oprima nueva mente la te...

Page 13: ...ansmisor que se usar Cuando se calibre con un transmisor se debe usar un aparato compatible Por ejemplo use un aparato para leer la temperatura para calibrar el ET6 usando un transmisor de temperatura...

Page 14: ...brazo 3 Deslice el estilete nuevo sobre el brazo del estilete hasta quede perfectamente ensamblado 4 Quite la cubierta del estilete Ins rtela en el sujetador para cubiertas de estilete 5 Oprima la tec...

Page 15: ...s a mal uso o negligencia Sintoma Causa Verificar Remedio La l nea trazada por el estilete La punta est demasiado Lije la punta humedezca es demasiado delgada o est afilada o el cartucho se sec o reem...

Page 16: ...u ce que la fl che de temps pointe vers l heure et le jour corrects sur le disque de trac voir Anatomie de l enregistreur la page 3 Avant d installer l enregistreur ET6 lire la notice en entier afin d...

Page 17: ...stallation sur panneau Contact de relais facultatif Enregistreur 2 thermocouples Adaptateur d alimentation secteur Trous d installation murale Trous d installation sur panneau Ensemble style de trac F...

Page 18: ...borne n gative gauche Serrer les vis du bornier pour fixer les fils du transmetteur R p ter l op ration avec le deuxi me bornier si le mod le deux styles est utilis Styles L hyst r sis un ph nom ne qu...

Page 19: ...lignoter L alarme sonore retentit galement si elle est en fonction commutateur DIP n 8 sur Arr t elle continue tre mise jusqu ce qu elle soit mise hors fonction ou que la cause de l alarme ait t limin...

Page 20: ...nage avec un transmetteur un talon compatible doit tre utilis Utiliser par exemple un talon de temp rature tr s pr cis pour talonner l enregistreur ET6 connect un transmetteur de temp rature 2 La proc...

Page 21: ...ksonWare Cette pointe repr sente le moment auquel le d faut d alimentation a commenc Les donn es enregistr es apr s cette pointe sont associ es une heure incorrecte Le logiciel DicksonWare calcule l h...

Page 22: ...Und Au dessous de la plage de V rifier la plage de temp rature l enregistreur talonnage d fectueux Milieux hostiles conditions Voir la proc dure d talonnage la difficiles ou passage du temps page 8 o...

Page 23: ...ie ne couvre pas l talonnage courant ou le rem placement des styles du disque de trac et de la pile La garantie ci dessus et le recours sont exclusifs et remplacent toute autre garantie explicite ou i...

Page 24: ...930 South Westwood Avenue Addison IL 60101 4997 Phone 630 543 3747 Fax 630 543 0498 E mail Dicksoncsr dicksonweb com www dicksonweb com MAN20 Revised 3 01 DICKSON www dicksonweb com...

Reviews: