23
Installation & Configuration
PC-9042 Portable Anechoic Chamber
Default Settings
The settings below are recommended speed settings for quiet and efficient movement, ultimately
custom settings will be used based on DUT characteristics. The values below are for the MCxx Series
Controllers.
Turntable Axis
Gear Ratio
90 :1
Motor Resolution
1.8 Degree
Step Mode
1/16th step or full step
Motor Current
2.0 Amps Max
Holding Current
0.44 Amps
Start Speed
1,000 steps per second @ 1/16th step resolution)
End Speed
Up to 12000 steps per second
Slope
Rate of acceleration from start to end speed
Home /Limit Sensor
Normally Open
Encoder
Programmable Quadrature Incremental - 800 CPR
MC3X specific controller commands with above Motor Current / Hold settings
Current Setting
P3,165,75,10
Step Mode:
H4
Gimbal Axis
Gear Ratio
90 :1
Motor Resolution
1.8 Degree
Step Mode
1/16th step or full step
Motor Current
2.0 Amps
Holding Current
0.44 Amps
Start Speed
1,000 steps per second @ 1/16th step resolution)
End Speed
Up to 12,000 steps per second
Slope
Rate of acceleration from start to end speed
Home /Limit Sensor
Normally Open
Encoder
Programmable Quadrature Incremental - 800 CPR
MC3X specific controller commands with above Motor Current / Hold settings
Current Setting
P3,165,75,10
Step Mode:
H4
Detailed Motor and Axis Specifications