Step by step instructions
LAC 142 / LAC 342 / GLU 142 / CHO 142 / TRI 142
Single measurement
Diaglobal GmbH · Köpenicker Straße 325 · 12555 Berlin · +49 (0)30 6576 2597 · [email protected] · www.diaglobal.de 2021-01
1. Insert capillary with 10 µL
sample into cuvette
2. Eject sample several times
with micropipetter into
cuvette
3. Screw cap on
Turn cuvette upside down
several times
5. Zero point adjustment: Insert
cuvette with sample (Fig. 3) into
photometer, zero point is stored
in memory
Remove cuvette after signal
tone
4. Switch photometer on with
ON/ENTER key
Wait for device check, confirm
with ON/ENTER
Select the required test,
confirm with ON/ENTER
6. Replace screw cap with
starter cap
ON
ENTER
ON
ENTER
M1 LAC
Sample 01
insert
9. Time is displayed, wait for
measured value
7. Turn cuvette upside down
several times
8. First press ON/ENTER
Then insert cuvette into
photometer
ON
ENTER
M2 LAC
Sample 01 00:30
Measuring
M1 LAC
Sample 02
insert or
ENTER