10
2021-01
7.2 Multi-point measurement with consideration of the sample blank value
and calculation of the endpoint
After measuring the sample blank value (=measurement 1) the colour reaction in
the cuvette is started. The course of the reaction is monitored by the instrument.
The endpoint is calculated using several absorbance values recorded at different
times.
Parameter: Glucose (GLU)
Measuring time: 2 minutes
8. Measurement
8.1 Switching the device on
Press the <ON/ENTER> key.
8.2 Self-test when switching on
When the device is switched on, a self-test of the digital and analogue circuitry is
conducted. The operational device check proceeds automatically after it is
switched on. It takes approx. 5 seconds, after which the unit is ready for
measuring
Note:
If it becomes obvious during the test that one of the device functions does not
correspond to the required settings, <SERVICE> will appear in the display.
In this case, switch the device off.
Please call Diaglobal GmbH service (Tel. +49 (0) 30 6576 2597) or contact your
specialist retailer.
8.3 Test selection
Press the <ON/ENTER> key.
The desired test is selected from the menu with the right or left arrow key:
HB-SLS - ERY - HCT - HB - GLU - ABS520 - ABS546
Pressing the right arrow key activates the next test while pressing the left arrow
key returns to the previous test. The selected test is shown in the upper right
corner of the display.
Confirm test selection with the <ON/ENTER> key.
8.4
Switching the device off
To switch the device off, press both arrow keys simultaneously.