Diaglobal DP 220 Operating Manual Download Page 11

11 

2021-01 

8.5

 

Integrated operational device checks 

 

 

Self-test when switching on 

Testing  of  the  digital  and  analogue  circuits  of  the  device  is  automatically 
performed by the device when it is switched on. 
Please see chapter 8, point 8.2. 

 
Differential measurements 

All measurements are based on differential measurements. I. e. after selecting the 
desired test, the device requests a zero measurement with a blank value cuvette. 
This creates a reference base to the measured value so that minor deviations can 
be compensated. 
 
Measurement range controls 

The  measurement  ranges  of  all  measurement  results  shown  in  the  display  are 
verified  by  an  integrated  measurement  range  control.  If  the  measurement  range 
is exceeded, an error is displayed. 
 
The  measurement  ranges  that  are  separately  defined  for  each  parameter  are 
documented on the respective package inserts as well as in this operating manual, 
chapter 9, Technical Data. 
 
Plausibility controls 

For multi-point measurements, the absorbance measured first forms the reference 
basis. The programme verifies the plausibility of the individual measured values. If 
specific requirements (e.g. E2 > E1 for ascending reactions) are not met, an error 
message is displayed. 

 
 
8.6     Notes on taking samples and carrying out measurements 

 
Errors in taking samples will always lead to incorrect measurement results. 
This  chapter  addresses  the  most  common  errors  that  can  occur  during  taking 
samples and measuring samples. 
 
1.

 

Before  measuring,  cuvettes  stored  in  a  refrigerator  must  be  brought  to  room 
temperature.  If  the  cuvettes  are  too  cold,  they  will  become  misty  with  water 
on  the  outer  wall  due  to  the  humidity,  which  will  lead  to  incorrect 
measurement results. 

 

2.

 

Never  touch  the  lower  part  of  the  cuvette  (where  the  liquid  is)  with  bare 
hands. If this should happen accidentally, clean the vials with a fluff-free cloth 
before use. Cleaning with a fluff-free cloth is recommended in any case. Even 
if  the  package  is  still  new  and  unopened.  Fingerprints  on  the  cuvette  lead  to 
incorrect measurement results. 

 
3.

 

If  the  blood  is  taken  from  the  heel  using  the  microvette  (neonatal  bilirubin), 
make sure that there is enough blood (approx. 60 µL) in the microvette, as 20 
µL  serum/plasma  is  needed  for  the  measurement.  Close  the  microvette  well 
after  taking  the  blood  sample  and  return  it  to  the  small  sample  tube  for 
centrifugation. 

 
 
 
 

Summary of Contents for DP 220

Page 1: ...1 2021 01 Gyn Photometer DP 220 Operating Manual Version 5 9 5 9 SI Edition 2021 01 ...

Page 2: ...gher an automatic test of the device function has also been integrated Therefore the Gyn Photometer complies with the requirements of the guidelines of the German Medical Association With the Gyn Photometer 4 clinical chemical parameters can be determined The device can be supplied with SI units of measurement on request see chapter 9 Technical Data table Measuring Ranges The kits and accessories ...

Page 3: ... Reagents parameter list 8 5 3 Control materials 8 5 4 Laboratory aids 8 6 Quality control according to the Guideline of the German Medical Association 9 7 Measuring procedure 9 7 1 End point measurement 9 7 2 Multipoint measurement taking the sample s blank value and calculation of the end point into account 10 8 Measurement 10 8 1 Switching the device on 10 8 2 Self test when switching on 10 8 3...

Page 4: ...nd serum plasma The Gyn Photometer fulfils the basic requirements of Appendix I of Directive 98 79 EC regarding in vitro diagnostics The conformity of the device with Directive 98 79 EC is confirmed by the use of the CE marking Manufacturer Diaglobal GmbH Innovationspark Wuhlheide Köpenicker Str 325 Haus 41 12555 Berlin Tel 49 0 30 6576 2597 Fax 49 0 30 6576 2517 E Mail info diaglobal de http www ...

Page 5: ...netic waves Operation on a horizontal surface Please observe the following instructions for use Insert a rechargeable battery or normal battery if the device is to be operated independently of a power supply or connect the photometer to a power supply unit Press the ON ENTER key Fig 1 to activate the internal device check which is automatically carried out by the device The device is then immediat...

Page 6: ... compartment cover Connect the battery to the push button contact and insert it into the device Replace the battery compartment cover and screw in the knurled screws Please note The Gyn Photometer can be operated using a power supply without the need to remove the rechargeable battery or the normal battery The rechargeable battery cannot be charged while it is installed A separate battery charger ...

Page 7: ...Mikrozid AF Liquid Bacillol plus 3 Kohrsolin or similar are recommended for cleaning the device and the surface Before cleaning the unit with a soft cloth and the decontaminating solution it must be switched off and the electrical power supply must be disconnected Make sure that no liquids get into the device There is no protection against penetrating liquids Code IP X0 The cuvette shaft must not ...

Page 8: ...tocrit control Control blood for accuracy and precision control of haematocrit and erythrocytes determination in blood in the normal range 5 x 1 mL GLU QS Glucose control 100 mg dL 3 x 4 mL 5 4 Laboratory aids Art no Description Contents LH 001 Blood lancets 500 LH 004 Capillaries 10 µL with ringmark 250 LH 006 Cuvette rack 1 LH 007 Micropipettor pipetting aid 1 LH 009 Cellulose swabs 500 LH 010 C...

Page 9: ...unts In agreement with the requirements of the German Medical Association a test of the device function see operating instructions chapter 8 5 is integrated in the Gyn Photometer therefor a daily test by means of a standard manual test German Medical Association part B chapter 2 1 5 paragraph 2 is not necessary The Gyn Photometer is suitable for the speedy detection of gestational diabetes and ful...

Page 10: ...after it is switched on It takes approx 5 seconds after which the unit is ready for measuring Note If it becomes obvious during the test that one of the device functions does not correspond to the required settings SERVICE will appear in the display In this case switch the device off Please call Diaglobal GmbH service Tel 49 0 30 6576 2597 or contact your specialist retailer 8 3 Test selection Pre...

Page 11: ... the individual measured values If specific requirements e g E2 E1 for ascending reactions are not met an error message is displayed 8 6 Notes on taking samples and carrying out measurements Errors in taking samples will always lead to incorrect measurement results This chapter addresses the most common errors that can occur during taking samples and measuring samples 1 Before measuring cuvettes s...

Page 12: ... 1 mm from the ring mark is sufficient to obtain a completely incorrect measurement result If the sample is above the black ring mark this will lead to incorrect positive measurement results A cellulose swab can be used to carefully soak up too much blood If the sample is below the black ring mark this will lead to incorrect negative measurement results In this case correction is hardly possible d...

Page 13: ... light influence Negligible Interface V24 9600 8 n 2 Power supply 6 V 12 V DC Current consumption max 250 mA Warm up time 0 min Interference suppression According to DIN VDE 0871 and DIN VDE 0875 Inaccuracy 0 5 at A 1 000 Relative photometric short time standard deviation 0 1 Measuring ranges DP 220 DP 220 SI Glucose 20 630 mg dL 1 1 35 mmol L Haemoglobin HiCN method 0 0 50 g dL 0 0 31 mmol L Haem...

Page 14: ...irements Part 2 6 Particular requirements In vitro diagnostic IVD medical equipment 9 EN 592 Instructions for use for in vitro diagnostic instruments for professional use National directives and recommendations Germany 10 Guidelines for Quality Assurance of Laboratory Examinations of the German Medical Association of 23 12 2019 Note on electromagnetic compatibility a The photometer meets the requi...

Page 15: ...ss ON ENTER key Wait for device check and confirm with ON ENTER 2 Select test Press arrow key until required test appears 3 Confirm required test Press ON ENTER 4 Switch off Press both arrow keys at the same time Note If SERVICE appears in the display after the device check the device has a defect In this case please contact our customer service at 49 0 30 6576 2597 ...

Page 16: ...ro point adjustment Insert an unprocessed cuvette from package into photometer Zero point is stored in memory 4 Switch photometer on with ON ENTER key Wait for device check confirm with ON ENTER Select the required test confirm with ON ENTER In regard to series measurement After zero point setting any number of additional samples can be measured ON ENTER ON ENTER HB SLS Insert blank Zero HB SLS 0 ...

Page 17: ...Zero point adjustment Insert cuvette with sample Fig 3 into photometer zero point is stored in memory Remove cuvette after signal tone 4 Switch photometer on with ON ENTER key Wait for device check confirm with ON ENTER Select the required test confirm with ON ENTER 6 Replace screw cap with starter cap ON ENTER ON ENTER M1 LAC Sample 01 insert 9 Time is displayed wait for measured value 7 Turn cuv...

Reviews: