diagenode Bioruptor plus User Manual Download Page 19

PAGE 19 

DIAGENODE 

BIORUPTOR

®

 

PLUS 

USER MANUAL

Diagenode Inc. North America

 / Phone: +1 862 209-4680 // Fax: +1 862 209-4681 // Mail: orders.na

@

diagenode.com

resulting in broader size distribution or larger peaks. Bath water should be pure distilled water, changed 
regularly.

 

2.

    

Sonication bath maintenance:

 The sonication bath metal surface is fragile and requires a careful maintenance. 

Use only soft sponge and distilled water to remove traces. Never use scratch scrub sponge since this would 
alter the ultrasonic wave emitter surface.

Supplementary Data

Please note that there are three main sources of variation in both peak base-pair size and distribution: 

1)   The physical process of DNA fragmentation might not be entirely random in AT- or GC- rich regions.

2)   The  analytical  process  to  determine  fragment  size  has  inherent  variances  (for  example,  gel  electrophoresis  and 

microfluidics-based platform). Therefore, fragment distributions and peak values, even from technical replicates, 
may not appear identical. If the sheared DNA sample will be resin or column purified or concentrated prior to analysis, 
please remember to take out an aliquot for use as control prior to that step. Column purification and concentration of 
the sheared DNA will generate a biased fragment distribution profile due to the inherent greater loss of the smaller 
DNA fragments.

3)   RNA contamination in genomic DNA preparation should be carefully removed using RNase-DNase free enzymatic 

digestion  since  they  might  generate  a  biased  fragment  distribution  profile  on  microfluidics-based  platform  (eg. 
Agilent Bioanalyzer) or alter sonication effiency.

Chromatin shearing

Critical points for chromatin shearing

•   Chromatin shearing efficiency varies on cell type. 

Each cell type might need additional protocol optimization.

•   The extent of cross-linking is critical for the efficient disruption of fixed cells and also affects DNA yield and average 

size  of  chromatin  fragments.  Over-cross-linked  chromatin  will  not  produce  small  fragments,  even  by  prolonged 
sonication. Fix cells for 8-10 min at RT, always stop the reaction by glycine and wash 2-3 times with ice cold PBS.

•   Cell  density  affects  the  sonication  efficiency.  Do  not  use  too  dense  cell  suspension.  Optimal  density  is  about 

1-3x10^6/100 μl of sonication buffer.

•   SDS is a key component of sonication buffer for chromatin shearing. Include 0.7-1% of SDS in your sonication buffer.

•   Fresh formaldehyde for fixation.

Shearing of chromatin from adherent cell lines

For the adherent cells, we recommend to first harvest cells by trypsinization and perform chromatin cross-linking in 
a cell suspension rather than on dishes as it results in a better reproducibility and consistency between experiments.

1.   Discard medium to remove dead cells and wash cells by adding cold PBS.

2.   Harvest cells by trypsinization.

3.   Transfer cells in a tube containing 10 ml PBS (RT) and centrifuge 5 minutes at 1.300 rpm. Keep the cell pellet and 

discard the supernatant. Wash the cells again in PBS.

Note:

 At this step, cells might be counted.

4.   Add PBS to a final volume of 500 μl for a 

maximum of 10x10^6 cells

 (for more cells, perform the fixation in a separate 

tube).

5.  Add formaldehyde to a final concentration of 1%, mix gently and incubate for 8-10 min at RT with rotation.

6.   Stop the cross-linking reaction by adding glycine to a final concentration 0.125 M and incubate for 5 min at RT with 

Summary of Contents for Bioruptor plus

Page 1: ...U S E R M A N U A L Version 1 1 Bioruptor Plus Cat No UCD 300 Bioruptor Sonication System ...

Page 2: ... expenses to return the repaired or replacement product This warranty covers only manufacturing defects and does not cover any damage caused by misuse lack of compliance to recommendations stated in the manual neglect accidents abrasion or exposure to extreme temperatures chemical solvents or acids We strongly recommend that maintenance or repairs of Diagenode s products are performed by our appro...

Page 3: ...tions 6 Getting to Know Your Bioruptor System 7 Bioruptor Components Overview 7 Sonication bath 7 Motorized lid 8 Metallic soundproof box 9 Power Supply Adapter 9 Tube holders 9 Equipment Installation 10 Installation overview 10 Installing the Bioruptor Plus 12 Controlling the Sonication 13 Tube Holders Tubes 15 Standard protocols 16 DNA Shearing 16 Chromatin Shearing 19 Bacteria Cell Disruption 2...

Page 4: ... to cool at least 20 minutes between runs Damage resulting from non compliance to manual instructions will void the warranty and shorten the lifespan of the machine Ultrasound Emitters can be damaged by tilting or jarring the machine Exercise care if moving water tank Validated tubes for the Bioruptor Plus DNA shearing 0 5 ml Bioruptor Microtubes for DNA Shearing Cat No WA 004 0500 Chromatin Shear...

Page 5: ... is called cavitation The bubble continues to absorb energy until it can no longer sustain itself and implodes This produces intense focused shearing forces that disperse and break biomolecules The fragmentation of DNA takes place as a consequence of this mechanical stress or shear With the Bioruptor the entire volume of water present in the sonication bath is exposed to ultrasound allowing all th...

Page 6: ...to be harmful to human health However we would recommend avoiding unnecessary exposure Diagenode recommends that pregnant women should not be exposed to 20 60 kHz wave lengths for a long period of time Bioruptor Technical Specifications Bioruptor Power Supply 115 V 4 2A US 230V 2 1A EU 50 60Hz Control unit dimensions 350 W x 260 D x 165 H mm Sonication unit dimensions sonication bath 175 W x 160 D...

Page 7: ...he apparatus is required after initial set up it is imperative to keep the tank at a right angle to the ground during the transport at all times Water level and quality The level of the water has been optimized and should always reach the red line sticker on the wall of the tank Use only distilled water to fill the tank Do not use deionized water Replacement stickers can be obtained from Diagenode...

Page 8: ... BioAcc Cool can be used in combination with the Standard Connector Kit continuous valve to guarantee the automatic temperature control of the sonication bath during the entire sonication process Figure 3 The water flow shouldn t exceed 500 ml min to allow optimal resonance The Water Cooler features two pumps and produces a regular water flow to maintain a constant water level in the tank An addit...

Page 9: ...h the power button located on the back Tube Holders Several sizes of tubes can be used with the Bioruptor Plus The minimum and maximum sample volume to be used with each container is given in the table below The 0 5 ml and 1 5 ml tubes can be simply closed and installed in the rotor For the sonication of larger volumes 10 ml 15 ml and 50 ml tubes a stopper with a metallic bar has to be used to ref...

Page 10: ...ider all hazards even though no particular hazards have been identified during installation Before starting installation work turn the main switch off back side of the control unit and secure the unit against being re energized No special tools are required Three 3 square meters are needed to set up the Bioruptor Devices are designed to be safe under the following conditions Water Connector kit fo...

Page 11: ...power inlet behind the power supply adapter shows the right voltage corresponding to your area Otherwise switch it using a narrow blade screwdriver This feature is not applicable for the Bioruptor Plus UCD 300 sold in the United States Attaching Cables When connecting cables always be sure pins are properly aligned Note the indexing pins on the control unit cables for precise mating alignment Keep...

Page 12: ...onication bath into the soundproof box 2 Place sonication bath in front of the soundproof box 5 Place the motor lid on the top of the sonication bath and connect it Make sure that the cable of the motor lid is connected as shown in the image triangle have to match 8 Connect the power supply adapter to the control unit with the power cable Power supply adapter is only needed outside the U S 9 Place...

Page 13: ...he display shows cycle 4 of 10 Bioruptor NGS will sonicate as shown CYCLE Num 10 Time ON 30sec Time OFF 30sec Press OK Modify Bath 1 ON OFF TIME ON TIME OFF CYCLE START Cycle number reached 4 cycles in this example Buttons and their functions Button A Pause restart after pausing Button B Press and hold this button during sonication to display T1 total on time per cycle and T2 total off time per cy...

Page 14: ...in a row recently without breaks Instrument is stored in a place exposed to direct sun Room temperature is too hot What to do Store the instrument in another place if the current one does not meet the installation specific requirements Allow for longer breaks between uses Place crushed ice in sonication bath to help the cool ultrasound source 2 During a run if the instrument stops and the followin...

Page 15: ... 10 ml tube holder Cat No UCD pack 10 Tube holders tubes Tube holders Available for 0 5 0 65 ml 1 5 ml 10 ml 15 ml and 50 ml To use adaptor tube unit remove the lower part of the microtube holder by turning counterclockwise Then place microtubes in the unit Attach the lower part to the upper part of the adaptor To guarantee homogeneity of chromatin shearing the tube holders should always be comple...

Page 16: ...ring For DNA shearing we highly recommend to use the tube holder for 0 5 0 65 ml tubes Cat No UCD pack 0 5 and the corresponding Bioruptor 0 5 ml Microtubes for DNA Shearing Cat No WA 004 0500 To use the tube holder remove the lower part by turning counterclockwise Then place microtubes into the unit Attach the lower part to the upper part of the adaptor To guarantee homogeneity of DNA shearing th...

Page 17: ...itions On Off times in sec Number of cycles 1250 bp 15 90 2 cycles 950 bp 15 90 4 cycles 750 bp 30 90 3 cycles 550 bp 30 90 5 cycles 400 bp 30 90 6 cycles 350 bp 30 90 8 cycles 300 bp 30 90 10 cycles 250 bp 30 90 15 cycles 200 bp 30 90 30 cycles 150 bp 30 30 70 cycles recommended to use a lab timer to set time precisely The protocol settings listed above are recommended guidelines and actual resul...

Page 18: ...NA preparation might generate a biased fragment distribution profile on microfluidics based platform eg Agilent Bioanalyzer or alter sonication effiency Therefore it is highly recommended to use only high quality DNA when sonicating DNA for next gen sequencing library preparation The DNA sample to be processed should be highly pure having an OD260 280 ratio of between 1 8 and 2 0 and should be as ...

Page 19: ...er sonication effiency Chromatin shearing Critical points for chromatin shearing Chromatin shearing efficiency varies on cell type Each cell type might need additional protocol optimization The extent of cross linking is critical for the efficient disruption of fixed cells and also affects DNA yield and average size of chromatin fragments Over cross linked chromatin will not produce small fragment...

Page 20: ...l lines Note Cells growing in suspension culture are known to be difficult to shear Nuclei extraction is recommended before sonication Do not use very dense cell suspension for sonication 1 Cross link chromatin with 1 fresh formaldehyde for 8 10 min at RT 2 Stop the cross linking reaction by adding glycine to the final concentration 0 125 M for 5 min at RT with gentle rotation 3 Wash cells 3 times...

Page 21: ...hed ice Power setting H position High Sonication cycle 30 sec ON 30 sec OFF Total sonication time 10 min for UCD200 300 15 min for Bioruptor XL Note Please note that additional optimization might be required depending on the bacterial strain and growth phase Gram positive bacteria are more resistant to sonication than Gram negative bacteria because of the rigid cell wall Cells in log phase are les...

Page 22: ...cells with intact membranes and discrimination from cells with damaged membranes Thus efficient cell disruption is observed after 5 10 minutes of sonication Figure 1 Effect of sonication on cell disruption The number of intact cells after sonication was determined by measuring optical density at 600 nm Optical density of the cell culture before sonication 0 min is arbitrarily set to 100 Sheared DN...

Page 23: ...ing step How long is the shearing Perform a time course for chromatin shearing It is possible to shear from 5 30 minutes If 30 interrupt sonication after every 10 minutes and centrifuge tubes briefly before proceeding with the remaining time What is the optimal cycle 30 seconds ON 30 seconds OFF What is the best temperature for shearing 4 C Make sure sonication bath is kept cool Once optimal condi...

Page 24: ...ard UCD 200 TM 1 5 ml UCD 200 TO 1 5 ml 15 ml UCD 200 TS 0 5 0 65 ml Bioruptor Plus UCD 300 TM 1 5 ml UCD 300 TO 1 5 ml 15 ml UCD 300 TS 0 5 0 65 ml Bioruptor Pico B01060001 Cooling System Water Cooler including continuous valve for Bioruptor BioAcc cool Consumables 1 5 ml TPX Microtubes M 50001 1 5 ml TPX Microtubes M 50050 15 ml TPX Microtubes M UN 15 0 5 ml Bioruptor Microtubes for DNA Shearing...

Reviews: