![diagenode Bioruptor plus User Manual Download Page 10](http://html.mh-extra.com/html/diagenode/bioruptor-plus/bioruptor-plus_user-manual_2492804010.webp)
PAGE 10
DIAGENODE
BIORUPTOR
®
PLUS
USER MANUAL
Europe Diagenode sa
/ CHU - Tour GIGA - B34 - 3e étage // Avenue de l’Hôpital, 1 // 4000 Liège (Sart Tilman) // Belgium // Phone: (+32) 4 364 20 50 // Mail: info
@
diagenode.com
• Indoor use
• Altitude up to 2,000 meters
• Operating external temperature 0°C to 25°C
• Maximum relative humidity 80%
• Transient overvoltage typically present on the
MAINS supply
• Degree of protection: IP20
• Power plug must be grounded
• POLLUTION DEGREE 2 (Normally only non-conductive
pollution occurs. However, occasionally a temporary
conductivity caused by condensation is expected)
• Never install this equipment in a place where
environmental conditions and warnings mentioned
above are infringed
Installation overview
Fig 5.
Schematic installation overview of the Bioruptor
®
Plus in combination with the Water Cooler.
Equipment Installation
The following pages contain information on installing your Bioruptor
®
Plus model. This equipment must only be installed
by personnel after reading this section. Consider all hazards even though no particular hazards have been identified during
installation. Before starting installation work, turn the main switch off (back side of the control unit) and secure the unit
against being re-energized. No special tools are required. Three (3) square meters are needed to set-up the Bioruptor
®
.
Devices are designed to be safe under the following conditions:
Water
Connector kit for
Water Cooler
Sonication bath
Motorized lid
Metallic soundproof box
Water Cooler
Power button
Stop button
Output selector switch
Start button
Power cord
Power cord
Level meter
Digital timer
0.5/0.65 & 1.5
ml
tube adaptor
10 ml, 15 ml
tube adaptor
50 ml
tube adaptor
Gentle ultrasound
Rotation
Flexible
format
Flexible
control
Cooling
Flexible format
Fits into current workflow
with standard tubes.
Scales with flexible sample
volume.
Rotation
Prevents contamination with
closed tubes.
Continuous rotation through
sonication bath guarantees
equal distribution of energy.
Flexible control
Easy to program.
Power range effectively
disrupts samples.
Gentle ultrasound
Gentle ultrasound method
preserves sample.
Cooling
Cooling system maintains
integrity of sensitive
samples.
1
2
3
4
5
1
2
3
4
5