9
I F U
Risk
Hazard
Risk Control Measure Information
User Risk
Device tilts.
If a lift is started with the user far outside the base of
support of the lift, or if a user sitting in a sling is moved
outside the base of support of the lift, it could lead to the
device tipping, and the user falling.
During a lift, the user must be over the base of support of
the lift during the entire lifting procedure. Never move the
user, or start a lift from, outside the base of support of the
device.
Material Risk
Components loosening or
breaking due to Corrosion.
Improper treatment of the lift could lead to components
corroding. Use the lift indoors, and do not store or use it in
wet or humid environments.
See sections Contraindications, Precautions and Warnings
and Maintenance for more information and detailed
information about use and storage environments.
Charging batteries
A tone sounding when using the lift indicates that the battery need recharging.
Charge the lift after use to ensure that the battery is always fully charged.
Lock the castor wheels when charging the battery.
Eva450/Eva600
1. Connect the charging cable to a power outlet and in the connection for charger cable.
The symbol for charging is shown on the display.
2. Check to ensure that the lamps on the control box light up. The green LED lamp indicates that
the charger is receiving power and the yellow LED lamp indicates that the battery is charging.
3. Charging stops automatically when the battery is fully charged.
Wall-mounted charger
1. Remove the battery pack from the lift and place it in the wall-mounted charger.
2. Check to ensure that the LED lamp on the front of the charger lights up.
Eva400
1. Connect the charger to the charger cable leading from the underside of the control box.
2. Connect the charger to a power outlet (100-240 V AC).
3. When the charger is connected, the control box lamp indicates a yellow light.
NOTE!
Before the lift is used for the first time, it must be charged for at least 4 hours.
For maximum battery life, charge batteries regularly. We recommend daily charging when the lift is
used daily.
The emergency stop must be deactivated during charging.
Eva450/Eva600: Battery information on the display
The battery discharging will be shown in four stages:
Battery state 1
: The battery is ok, no need for charging (100 - 50 %).
Battery state 2
: Battery needs charging. (50 - 25 %)
Battery state 3:
Battery needs charging. (Less than 25 %) A tone sounds when a button is pressed in
this battery state.
Battery state 4
: The battery needs charging. (17V or lower) At this stage some of the functionality
of the lift is lost.At this battery stage it is only possible to drive the lift arm down. Furthermore an
audio signal will sound when a control button is activated. The symbol will switch between the two
pictures for 10 seconds.
The battery symbol is shown when the control box is active until power down (2 minutes after use).
It is not possible to use other batttery types than BAJ1/BAJ2.
The battery level is measured via voltage. This means that it is possible to experience e.g. that the
battery switches from state 1 to state 2 and back to state 1.
LED
indication of
charging
LED indica-
tion of hand
control
activation
Button for
electrical
emergency
lowering
Charger
Eva400