250
I F U
Table 206 – Recommended separation distances between portable and mobile
RF communications equipment and the
EQUIPMENT
or
SYSTEM
-
for
EQUIPMENT
and
SYSTEMS
that are not
LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the Eva400EM/EE.
The Eva400EM/EE is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Eva400EM/EE can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Eva400EM/EE as
recommended below, according to the maximum output power of the communications equipment
Separation distance according to frequency of transmitter
m
Rated maximum
output of transmitter
W
150 kHz to 80 MHz
P
d
17
,1
=
80 MHz to 800 MHz
P
d
17
,1
=
800 MHz to 2,5 GHz
P
d
34
,
2
=
0,01
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,69
3,69
7,38
100
11,67
11,67
23,33
For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.