Windows network configuration
Please enter the Control Panel, go to the Network Connections and right-click on the Port, where your Photron camera is connected
to. Go to Properties and select the TCP/IPv4 protocol. Depending on the model the default IP address is usually:
camera IP: 192.168.0.10 / set computer to 192.168.0.XXX. If a connection cannot be established, please try:
camera IP: 192.168.10.0 / set computer to 192.168.10.XXX
The subnet is always 255.255.255.0. This is the default for the currently available cameras. Please refer to the Photron camera
manual for further information about the default IP address. Please enter
Camera IP reset (factory default)
In case you have troubles connecting to the camera, you can try to set it to factory default by the following routine:
Press and hold the RESET switch at the camera’s backside.
All of the LEDs on the camera's back sidelight, then they turn off sequentially from right to left. In the end, all LEDs
together blink twice, then stay on. This takes several seconds, hold the button pressed all the time.
Then reboot the camera. The IP address is reset to the factory setting.
Photron FASTCAM Viewer
For checking the network connection (and changing the camera settings), you can use the previously installed Photron FASTCAM
Viewer to get a live picture. Please enter the IP address mask as shown above, then click Detect, an auto-search will be performed.
Alternatively, click File / Camera open.
47