Additional information
IPROXX 2
The following standards and norms were used in accordance with EN 60601-1, IEC 60601-1,
Edition 3.1,
EN 60601-1, Section 7.8.1
Colours of indicating lights
EN 60601-1, Section 8.10
Components and wiring
EN 60601-1, Section 8.10.4.1
Operating voltage / Input voltage
EN 60601-1, Section 9.3
Hazards associated with surfaces, corners and edges
EN 60601-1, Section 11
Temperature
EN 60601-1, Section 13
Hazardous situation and fault conditions:
15W limited supply circuit
EN 60601-1, Section 15.4.7.1
Mechanical strength
EN 60601-1, Section 15.4.7.2
Accidental operation
EN 60601-1, Section 16.6
Leakage current: Touch current <100µA
EN 60601-1, Section 17
EMC
The following standards and norms were used in accordance with EN 60601-2-52, IEC 60601-2-52,
third Edition, (Particular requirements for the safety and essential performance of medical beds), the
following standards have been used
EN 60601-2-52, Section 201.3.8
Applied Part B
EN 60601-2-52, Section 201.9.2.2.5
Hold to run controls / momentary contact switch
EN 60601-2-52, Section 201.9.2.3.1
Unintended movement:
Means to deactivate IPROXX 2 SM, IPROXX 2 SM+,
IPROXX 2 SMP
EN 60601-2-52, Section 201.11.6.5.101
Ingress of water: min IPX4
EN 60601-2-52, Section 201.11.1.1
Temperature
EN 60601-2-52, Section 201.15.3.4.1
Falling test
EN 60601-2-52, Section 201.15.4.4
In
dicator lights “Ready for Normal Use” not required
EN 60601-2-52, Section BB.3.3.3
Dimensions for handles
EN 60601-2-52, Section BB.3.4.1
Operating forces