Troubleshooting
IPROXX 2
28
74828(c)
5.1 Self-test of the IPROXX 2 handset
The IPROXX 2 handset carries out a self-test in the following situations:
•
When switching from locked mode to programming mode or to patient mode
•
When switching between the patient mode and the nursing staff mode
•
When the handset is started
If one or more keys are pressed or jammed during the self-test, the handset will switch to an error
mode and the error will be displayed as follows:
•
The corresponding locking LEDs for the affected key pairs flash yellow
The handset must be switched off before it can be switched on and operated again. Pull the connec-
tor, wait 30 seconds, and then reconnect the connector plug.
The handset will also switch to error mode whenever a movement key is pressed for more than 60
seconds on the handset or in the app.
5.2 Malfunctioning of the permanent connection between the IPROXX 2 and iOS
smart phone
If the permanent connection between the smart phone and the IPROXX 2 handset is interrupted, the
app displays the message
Pairing error
. You should delete the connection data on the smart phone
to remedy this error.