ECO BETADRIVE ECO
Note: Operating information
81417 3.0
19
6.
Note: Operating information
The installation instructions do not contain all information required for the safe operation of the end
product. They only describe the installation and operation of the drive as components (as defined in
the Machinery Directive as a "partially assembled piece of machinery").
Power-on time / intermittent operations
NOTICE
The ECO BETADRIVE ECO has been designed for intermittent operations. Intermittent opera-
tion is an operational mode where the drive must pause after a specified maximum period of
operation (power-on time). The drive can overheat when these pauses are not maintained.
The ECO BETADRIVE ECO must operate with an intermittent duty of 1 min/ 8 min. This means
that after the unit is operated with its rated load for up to one minutes it must then be paused for
8 minutes.
Shutting off the drive
In order to shut off the drive, unplug the mains power plug and the battery plug (when present)! The
power plug must always be accessible during operations so that emergency shut-off is possible.
Avoiding cable damage
Be sure that your operating instructions inform the user about the possible cable risks.
NOTICE
The cables (in particular the electrical connection for the drive and the power supply cable for the
drive system) must not be squashed or crushed. In order to prevent damage to the drive, no
mechanical strain should be placed on the cables.