DeVilbiss Healthcare drive Artemis II Instructions For Use Manual Download Page 36

36

14. 

ELECTROMAGNETIC COMPATIBILITY (EMC)

The control unit has been designed to meet the EMC requirements of EN 60601-1-

2 however it may still be affected by or emit harmful radio frequency (RF) energy. 

The RF emissions from the control unit are very low and are not likely to cause any 

interference to nearby electronic equipment, however interference to sensitive 

equipment is still possible. Likewise if the immunity limits of the control unit are 

exceeded the system may be seen to operate abnormally.
If the control unit or any alternative equipment is found to be operating abnormally 

turn off the piece of equipment that is believed to be causing the interference 

(if possible) to identify the source of the RF energy. Once identified mitigation 

measures are to be taken, such as the separation distances being increased and/

or the device(s) being re-orientated.
The dynamic system is to be installed and put into service according to the 

information provided within this section to ensure of reliable operation, however if 

the control unit continues to operate abnormally turn off at the mains supply and 

contact Drive DeVilbiss Healthcare Ltd. or your local distributor (see section 2).

•  Portable RF communications equipment (e.g. mobile/cordless 

phones) should be used no closer than 30 cm (12”) to the 

control unit or its mains cable, otherwise degradation of the 

performance of this equipment could result.

•  Use of accessories other than those specified or provided 

by Drive DeVilbiss Healthcare could result in increased 

electromagnetic emissions or decreased electromagnetic 

immunity of the control unit and result in improper operation.

•  The control unit should not be used adjacent to or stacked with 

other equipment where possible, if adjacent or stacked use is 

necessary the control unit should be observed to verify normal 

operation in the configuration in which it is to be used. 

Warning

*Not suitable for use in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous 

oxide (Not AP or APG equipment)
**If the system and its components are serviced and maintained in accordance with the information 

detailed in section 12 of these instructions for use then the system can be expected to provide in 

excess of the 3 years of service.

*** IP21 = water ingress protection from condensation only.

Summary of Contents for drive Artemis II

Page 1: ...Artemis II Dynamic Mattress and Cushions Instructions for use...

Page 2: ...button FIG 4 and pulling away FIG 5 Switch off the control unit CPR mattress only Rapid deflation of the mattress may be required for emergency treatment or system deflation The CPR dial is located a...

Page 3: ...es if power is lost 7 MAX INFLATE MODE COMFORT CONTROL Opens comfort control menu 1 Press MAX INFLATE MODE COMFORT CONTROL to open Comfort Control Menu 2 Press down MAX INFLATE MODE COMFORT CONTROL MU...

Page 4: ...ng Mode 20 9 5 2 2 Patient Weight Pressure 21 9 5 3 Safety Functions 21 9 5 3 1 Function Lock 22 9 5 3 2 Mute 22 9 5 4 System Functions 22 9 5 4 1 Mode Selection 22 9 5 4 2 Max Inflate 23 9 5 4 3 Comf...

Page 5: ...DI and serial numbers can be found on the back of the control unit and inside the mattress cushion For Service Support outside the UK Northern Ireland please contact the local distribution company fro...

Page 6: ...n overall programme of care when active load distribution through mechanical means is required 3 3 Intended Use The intended use of the mattress is to support the weight of the patient as identified w...

Page 7: ...e to a pre determined value but the comfort level can be adjusted by manually adjusting the cell pressure up or down Should a fault occur such as a power failure or loss of pressure an audio visual al...

Page 8: ...lity of the patient to the mattress size Patient with burns Unauthorised people with access to the controls Patients who have reduced capacity and are agitated and or restless Small adults children Su...

Page 9: ...aters away from the support surface Follow all manufacturers instructions and warnings It is advised that a full fire risk assessment is carried out prior to using this equipment In case of fire exit...

Page 10: ...n fire Control unit functions must be locked out when a patient is left unattended If children adults who lack capacity or even pets pose a potential risk of intentional or unintentional tampering wit...

Page 11: ...torage ensure it has been cleaned and disinfected in line with the local infection control policy and or as defined in section 10 of these instructions for use Do not remove the mattress cushion from...

Page 12: ...ot dry clean Tumble dry on low heat Zip location Keep out of direct sunlight Max patient weight Medical device MD SN Serial Number No Smoking Temperature limit Class II Electrical Device The user is p...

Page 13: ...parts of the device that come into contact with the patient in order to carry out its intended function refer to section 16 2 Type BF Applied parts which are electrically isolated from earth and other...

Page 14: ...ication If the system has come from a storage transport temperature environment near to the minimum or maximum values stated allow the cushion mattress and control unit to adjust to room temperature f...

Page 15: ...raps For profiling beds it is essential that the straps are secured around the movable sections of the bed frame Damage will be incurred when profiled if secured to fixed parts of the frame To avoid a...

Page 16: ...ttress specification could lead to an injury The mains plug is the disconnect device for the means of isolating the control unit from the mains supply the plug must be accessible at all times Ensure t...

Page 17: ...oisture could damage the electrical system and pose an electrical fire risk Ensure the cushion is used with a compatibly sized chair incorrect combinations can lead to entrapments falls hazards Incomp...

Page 18: ...menu 8 Mode select 9 3 Control Interface No Symbol Description 9 Enters maximum inflation 10 Returns to Home screen No Symbol Description 1 Turns system on off 2 Mutes the audible signal for a 20 minu...

Page 19: ...Cycle Time UP DOWN KEY 4 Extended Functions Page 2 2 Comfort Control Menu Press Enter Button to Enable Selection Home Enter Key Mode Key Comfort Control Key Home Enter Key Mode Key Home Screen Comfort...

Page 20: ...r 45 minutes see section 11 To turn the system off hold the button down for 2 seconds until the screen extinguishes Once a patient is positioned on the mattress the system will determine the patient s...

Page 21: ...t indicator is not to be used for medical purposes it is to act as information only If the inclusion of the weight indicator could introduce a hazard it can be hidden see 9 5 4 3 Warning Once the syst...

Page 22: ...ng When pressed a menu appears offering 3 different settings 9 5 4 System Functions 9 5 4 1 Mode Selection 1 Alternating 2 Constant Low Pressure 3 Pulsation Use the up down cursor key followed by the...

Page 23: ...fety Before changing the pressure or cycle time clinical judgement is required from frequent monitoring and repositioning of the patient 9 5 4 3 Comfort Control 9 5 4 4 Comfort Level When pressed a co...

Page 24: ...3 Auto Lock a On b Off When pressed a menu appears offering 2 different settings 4 Weight Unit a kg b lbs By selecting this function the cycle time of the alternation sequence can be manually altered...

Page 25: ...activate the egress alert If active and the occupant gets out of bed an audible signal sounds and a warning screen illuminates for approximately 1 minute followed by a no patient indication To cancel...

Page 26: ...t your service engineer Service Due Service due contact maintenance provider as soon as possible 9 6 2 Cushion Mode The seat cushion can be operated with the Artemis II pump the system will be automat...

Page 27: ...lling away from the control unit 2 9 8 Support Surface Disconnection and Power Cuts 1 2 Switch off the control unit if still operational Disconnect from the power supply Note on the mattress there is...

Page 28: ...cleaning must be carried out in accordance with your local infection control policy or regulatory body Warning Always disconnect the support surface and bed frame from the main power supply prior to...

Page 29: ...de Any evidence of strike through and or cover damage will require a new cover to be fitted to the mattress Warning The cover must not be used if strike through is evident Risk of cross infection Freq...

Page 30: ...pproved service engineer 1 Turn off the control unit to silence the alert and turn off the mains supply Note the mute button does not silence the power failure indication 2 Check the mains cable is fu...

Page 31: ...n on again to clear the indicator If a high pressure indicator is still evident turn off at the mains and contact your approved service provider Mattress Disconnection Audible signal sounds Screen sho...

Page 32: ...imum Drive DeVilbiss Healthcare Ltd also recommends that the carer performs frequent visual and operational inspections If there are any signs of damage or the system is not performing as it should wi...

Page 33: ...n alternatively follow local recycling and disposal policies The control unit used with your system is not to be disposed of in general municipal waste as it is to be considered as Waste Electrical an...

Page 34: ...0 15 or 20 minutes Pressure Range Alternating Mode 13 70 2mmHg Constant Low Pressure Mode 10 40 2mmHg Pulsation Mode 30 30 of CLP setting Max Inflate 70 2mmHg Control Unit Dimensions L 180mm x W 390mm...

Page 35: ...0 15 or 20 minutes Pressure Range Alternating Mode 40 90 2mmHg Constant Low Pressure Mode 10 40 2mmHg Pulsation Mode 30 30 of CLP setting Max Inflate 90 2mmHg Control Unit Dimensions L 180mm x W 390mm...

Page 36: ...to operate abnormally turn off at the mains supply and contact Drive DeVilbiss Healthcare Ltd or your local distributor see section 2 Portable RF communications equipment e g mobile cordless phones s...

Page 37: ...vel Compliance level Electrostatic discharge ESD IEC 61000 4 2 8 kV contact 15 kV air 8 kV contact 15 kV air Electrical fast transient burst IEC 61000 4 4 2 kV for power supply lines 2kV for power sup...

Page 38: ...bed frames which have a platform of approximately 2000mm x 900mm however the choice of side rail used can determine suitable compatibility or not Drive DeVilbiss Healthcare bedframes reference suitabl...

Page 39: ...iled by or attributes to acts of God f damaged caused by failure to follow instructions or g otherwise used in a manner inconsistent with any instructions provided by Drive DeVilbiss Healthcare Ltd Th...

Page 40: ...vedevilbiss co uk www drivedevilbiss co uk Drive DeVilbiss Healthcare Ltd Sidhil Industrial Estate Holmfield Halifax HX2 9TN West Yorkshire United Kingdom A member of the Drive DeVilbiss Group of Comp...

Reviews: