23
3) Pulsation: Creates tissue stimulus by alternately increasing and decreasing the
cell pressure of the constant low pressure mode by 30% in each direction.
When using the pulsation setting, to return to alternating it is necessary to
manually re-select alternating from the therapy menu, it will not automatically
default back.
Weight(kg): 40■■■■■■□□□□□□□□□□□□□□267
Mode: Pulsation
Pressure Setting: 7 - 13 mmHg
Cycle Time: 10 minutes
9.5.4.2 Max inflate
Inflates the cells to maximum pressure to provide a stable, static support surface.
The system will automatically revert back to alternation mode after 20 minutes for
patient safety.
Before changing the pressure or cycle time, clinical judgement is required from
frequent monitoring and repositioning of the patient.
9.5.4.3 Comfort Control
9.5.4.4 Comfort Level
When pressed a comfort level control menu appears:
1: Comfortable Level
a: Firm
b:
Standard
c:
Low
Use the up/down cursor key followed by the selection key to initiate the chosen
setting.
By selecting this function the softness / firmness of the mattress can be manually
altered, dependent on the patient’s requirements. The pump defaults to medium.
Firm =
+ 20% mmHg
Medium =
Automated Pressure Setting
Soft =
– 20% mmHg
CLP mode will not automatically default back to alternating mode.
Alternating mode must be reselected manually by the user.
Warning
The control unit will not return to alternating mode unless manually selected by the
user.