![DeVilbiss Healthcare drive Artemis II Instructions For Use Manual Download Page 30](http://html1.mh-extra.com/html/devilbiss-healthcare/drive-artemis-ii/drive-artemis-ii_instructions-for-use-manual_2489318030.webp)
30
10.3
Alternative Cover Cleaning Instructions
Alternatively disinfection of the cover may be achieved by laundering as follows:
• Remove cover.
• Machine wash at 71°C for not less than 3 minutes or 65°C for not less than 10
minutes. Heavily soiled items should also have a pre-wash/sluice cycle.
• Allow covers to fully dry before use.
(Refer to the Department of Health document HTM 01-04 for further details).
Warning
The control unit is not to be opened – risk of electrocution.
11. TROUBLESHOOTING
Symptoms
Indications
Actions
Power Failure
•
Amber ‘power
failure’ light flashes.
•
Audible signal
sounds.
•
Screen extinguishes.
If mains plug, cable or outer casing is visibly
damaged turn off at the mains and contact your
approved service engineer.
1.
Turn off the control unit to silence the alert and
turn off the mains supply (Note, the mute button
does not silence the power failure indication).
2. Check the mains cable is fully connected to the
control unit and plugged into a wall socket.
3. Switch on at the wall (to ensure the socket is
working, plug in a fused device that is known to
work).
4. Turn on the control unit.
If control unit still fails to operate turn off at the mains
and contact your approved service provider.
Service Due
•
Audible signal
sounds.
•
Screen shows:
Service due alert will present itself after 8760hours
of service when the control unit is turned on. The
alert will disperse and the system can be used as
normal. To reset the service due alert, contact your
service provider to arrange the necessary service &
maintenance to be performed.
**** Service Due ****
Service due, contact maintenance