Air Supply Hose

The air supply hose is self sealing when disconnected from the instrument. The hose
connects to the instrument with a quick release bayonet connection.

To attach the hose:

  Grip the hose connector and push it on to the instrument spigot.

  Twist the connector clockwise. The connector will come to a stop - the hose is now

connected.

  Pull gently on the hose to check that the hose is securely connected to the

instrument.

To remove the hose:

  Hold the instrument and grip the hose connector.

  Twist the hose connector anti clockwise, pushing the hose connector towards the

instrument.

Refer to the diagram.

Optional Accessories

Air Supply Hoses

Type

3m length

4m length

5m length

BOC MA7

5300

5310

5320

A. O.

5330

5340

5350

AGA

5360

5370

5380

Draeger

5390

5400

5410

Summary of Contents for DPX-170

Page 1: ...Operating Instructions Pneumatic Oscillating Saw International English Sternum Saw DPX 170...

Page 2: ......

Page 3: ...Safety Instructions 2 Product Description 2 Air Supply 2 Reprocessing Instructions 3 Air Supply Hose 6 Controls 7 Blade Clamp Operation 8 Conditions for Transport and Storage 8 Explanation of Symbols...

Page 4: ...AFE position when changing accessories or when not in use Always use eye protection when cutting to prevent injury from flying debris Do not allow loose articles to be caught by the moving parts of th...

Page 5: ...rried out where automatic washer disinfection is not available It should be conducted in a dedicated area by trained personnel wearing protective clothing e g gloves waterproof apron and goggles or vi...

Page 6: ...lude Pressurized cold water rinse maximum 35 C Hot water wash minimum 55 C using a neutral pH enzymatic detergent Warm water rinse Disinfection rinse minimum 80 C for 1 minutes Drying cycle 4 Remove d...

Page 7: ...s in accordance with EN868 is recommended to preserve sterility The material should present a barrier to micro organisms and particulate contamination Additional Information Automated cleaning was val...

Page 8: ...The connector will come to a stop the hose is now connected Pull gently on the hose to check that the hose is securely connected to the instrument To remove the hose Hold the instrument and grip the h...

Page 9: ...13 22 S81 104 0 5 1 0 1 5 42 51 S81 105 0 5 1 0 1 8 23 32 S81 106 0 5 1 0 1 4 15 25 For full details of suitable saw blades refer to Stericut data sheets Controls Lever The lever control has a progre...

Page 10: ...lined hole in the blade over the spline on the shaft Replace the clamp nut and tighten Removal Using the spanners provided remove the clamp nut on the end of the oscillating shaft in an anti clockwise...

Page 11: ...ufficient lubrication Refer to Lubrication section of manual Insufficient air supply pressure and or flow rate Check air supply pressure at regulator bottled air only Check hose for possible restricti...

Page 12: ...nd send to the address below Please ensure that the instrument has been properly decontaminated and sterilised Guarantee Liability De Soutter Medical guarantees all instruments attachments and accesso...

Reviews: