Fault Finding
Fault:
Possible Cause:
Remedy:
Instrument does not
run.
Air supply faulty
Check air supply
Check hose connections
Lever Mode
Selector set to
Safety
Rotate Lever Mode Selector to run
position
Instrument runs
slowly, or seems to
lack power
Insufficient
lubrication
Refer to Lubrication section of manual
Insufficient air supply
pressure and/or flow
rate
Check air supply pressure at regulator
(bottled air only).
Check hose for possible restriction.
Regulator
malfunctioning.
Check air supply regulator.
Blade will not fit onto
oscillating shaft
Debris inside blade
clamp
Clean with small brush
Blade is damaged
Replace the blade.
Do Not
force a damaged blade onto
the oscillating shaft
Instrument seems to
cut slowly
Worn accessory
Replace accessory.
If any problem persists then contact De Soutter Medical Ltd.