Warnings & Safety Rules
Failure to follow these instructions may result in serious injury to the patient
or operating staff.
Do not attempt to use the equipment until all the instructions have been
studied and understood.
Never
permit untrained personnel to use this instrument system.
Always
operate the instrument at the correct air pressure as specified in these
instructions.
Always
ensure that the hose is securely connected prior to use.
Always
inspect all equipment and accessories before use. Do not use suspect,
damaged or worn equipment.
Always
ensure accessories are correctly connected to the instrument before use.
Always
set the instrument safety catch to the
SAFE
position when changing
accessories or when not in use.
Always
use eye protection when cutting to prevent injury from flying debris.
Do
not allow loose articles to be caught by the moving parts of the instrument.
Never
drop the instrument or it’s accessories; always handle with extreme care.
Always
allow the instrument to stop before removing from the surgical site.
Only
clean and sterilise the instrument and accessories as directed in these
instructions.
Do not
immerse the instrument in fluids.
Always
use Stericut or De Soutter Medical approved accessories.
Product Description
The DPX-170 is a pneumatic surgical oscillating saw.
Air Supply
Use only 99.97% pure, dry compressed air or nitrogen filtered to 5 microns.
Lower pressures may be set for lower speed requirements.
Operating Pressure:
6 bar (87psi)
Flow Rate:
3.3l/s (7 cu.ft./min)
Never
exceed 7 bar (100psi) pressure.