Desoutter DPX-170 Operating Instructions Manual Download Page 12

Repair & Service Information

All powered instruments `should be periodically checked and cleaned. Annual
servicing is recommended for normal use. Due to the specialist techniques used in
the manufacture and maintenance of De Soutter instruments, user servicing is not
possible. For service and repair please contact your nearest De Soutter Medical
Authorised Service Centre.

To return an instrument for repair:

  Record the serial number of the instrument being returned. Enclose a brief

statement describing the reason for returning the instrument.

  Enclose the purchase order number for the instrument if warranty is being

claimed. It will be helpful to include a contact name.

  Pack the instrument securely and send to the address below.

Please ensure that the instrument has been properly decontaminated and sterilised.

Guarantee & Liability

De Soutter Medical guarantees all instruments, attachments and accessories to be
free from defects in material and workmanship for one year from the date of
purchase. De Soutter Medical is not liable by warranty or otherwise in the case of
any of the following:

  Abuse, misuse or use in other than a surgical environment;

  Disassembly, alteration or unauthorised repair;

  If the product has not been used in a reasonable manner and in full

compliance with the written instructions.

This guarantee does not affect your Statutory Rights in accordance with
1999/44/EEC

De Soutter Medical Ltd.

River Park, Billet Lane,

Berkhamsted, Herts. HP4 1HL, United Kingdom.

Email: [email protected]        Website: http://www.de-soutter.com

Tel: +44 (0) 1442 860300        Fax: +44 (0) 1442 860333

Ref: 1167en – 4.1

Summary of Contents for DPX-170

Page 1: ...Operating Instructions Pneumatic Oscillating Saw International English Sternum Saw DPX 170...

Page 2: ......

Page 3: ...Safety Instructions 2 Product Description 2 Air Supply 2 Reprocessing Instructions 3 Air Supply Hose 6 Controls 7 Blade Clamp Operation 8 Conditions for Transport and Storage 8 Explanation of Symbols...

Page 4: ...AFE position when changing accessories or when not in use Always use eye protection when cutting to prevent injury from flying debris Do not allow loose articles to be caught by the moving parts of th...

Page 5: ...rried out where automatic washer disinfection is not available It should be conducted in a dedicated area by trained personnel wearing protective clothing e g gloves waterproof apron and goggles or vi...

Page 6: ...lude Pressurized cold water rinse maximum 35 C Hot water wash minimum 55 C using a neutral pH enzymatic detergent Warm water rinse Disinfection rinse minimum 80 C for 1 minutes Drying cycle 4 Remove d...

Page 7: ...s in accordance with EN868 is recommended to preserve sterility The material should present a barrier to micro organisms and particulate contamination Additional Information Automated cleaning was val...

Page 8: ...The connector will come to a stop the hose is now connected Pull gently on the hose to check that the hose is securely connected to the instrument To remove the hose Hold the instrument and grip the h...

Page 9: ...13 22 S81 104 0 5 1 0 1 5 42 51 S81 105 0 5 1 0 1 8 23 32 S81 106 0 5 1 0 1 4 15 25 For full details of suitable saw blades refer to Stericut data sheets Controls Lever The lever control has a progre...

Page 10: ...lined hole in the blade over the spline on the shaft Replace the clamp nut and tighten Removal Using the spanners provided remove the clamp nut on the end of the oscillating shaft in an anti clockwise...

Page 11: ...ufficient lubrication Refer to Lubrication section of manual Insufficient air supply pressure and or flow rate Check air supply pressure at regulator bottled air only Check hose for possible restricti...

Page 12: ...nd send to the address below Please ensure that the instrument has been properly decontaminated and sterilised Guarantee Liability De Soutter Medical guarantees all instruments attachments and accesso...

Reviews: