Maintenance

The Digital Surface Resistance Meter requires little 

maintenance. There are no user serviceable parts. If the 

meter requires service beyond cleaning the electrodes 

or replacing the batteries, please contact the 

Statguard 

Flooring Customer Service

.

Battery Replacement

Replace the batteries once the battery life indictator 

is empty. Open the compartment located on the back 

of the meter to replace the batteries. The meter uses 

four AA alkaline batteries. Ensure that the batteries’ 

polarities are oriented in the correct fashion to avoid any 

possible circuit damage.

Cleaning the Digital Surface Resistance Meter

The area surrounding the test jacks at the top end of the 

meter should be wiped with a clean, isopropanol-alcohol 

moistened cloth to remove skin oils that will accumulate 

and affect the meter’s accuracy at high resistances. 

The frequency of cleaning will depend on usage. Desco 

recommends cleaning this area once a month. Cable 

jackets should also be cleaned in this fashion.

Cleaning the 5 lbs. Electrodes

Per ANSI/ESD S4.1 “Clean the electrodes with a 

minimum 70% isopropanol-water solution. Make sure 

the 5 pound electrodes’ conductive pads are dry prior to 

use.” 
See specific product test standards for test lab 

specimen cleaning instructions. Per ANSI/ESD S4.1 

Worksurfaces “The test specimens and electrodes 

shall be cleaned twice with a minimum 70% 

isopropanol-water solution using a clean, low-linting 

cloth each time.” (Note: The item should then be 

conditioned for 72 hours minimum)

Figure 13. Measuring the Resistance Point-to-Point (Rtt) 

of floor materials

Specifications

Resistance Ranges

1 x 10

3

 to 1 x 10

6

 ohms @ 10 

Volts, complies with ANSI/

ESD S4.1
1 x 10

6

 to 1 x 10

12

 ohms @ 

100 Volts, complies with 

ANSI/ESD S4.1

Resistance Accuracy

(@ 23ºC, 40% R.H.)

Resistance measurements 

within ±10% (±20% accuracy 

for 5 x 10

3

 ohms and lower 

and 1 x 10

12

 ohms and 

greater), complies with 

ANSI/ESD S4.1. Under load 

voltages of 10 Volts ±5% 

and 100 Volts ±5% exceeds 

requirements of ANSI/ESD 

S4.1

Temperature Accuracy ±10%
Relative Humidity 

Accuracy

±10 integers

Power Supply

4 AA alkaline batteries

Battery Life

Approximately 1,500 

measurements

Display

OLED, 2.7" diagonal, 128 x 

64 pixel resolution

Memory Capacity

100 measurements

Operating Temperature 41ºF to 85ºF (5ºC to 30ºC)
Environmental 

Requirements

Indoor use only at altitudes 

less than 6500 ft. (2 km)
Maximum relative humidity of 

80% up to 85°F (30°C) 

Dimensions

(meter)

3.94" x 8.27" x 1.26"

(100 mm x 210 mm x 32 mm)

Dimensions

(carrying case)

15.0" x 14.5" x 3.5"

(381 mm x 368 mm x 89 mm)

Weight

(meter with batteries)

0.9 lbs (0.4 kg)

Weight

(populated carrying 

case)

13.0 lbs (5.9 kg)

Country of Origin

United States of America

Limited Warranty, Warranty Exclusions, Limit of 

Liability and RMA Request Instructions

See the Statguard Flooring Warranty - 

StatguardFlooring.com/Limited-Warranty.aspx

Summary of Contents for Statguard Flooring D19290

Page 1: ...STM97 1 Floor Materials and Footwear Resistance Measurement in Combination with a Person A Compliance Verification Plan shall be established to ensure the Organization s fulfillment of the technical requirements of the ESD Control Program Plan Process monitoring measurements shall be conducted in accordance with a Compliance Verification Plan that identifies the technical requirements to be verifi...

Page 2: ...O INDUSTRIES INC Employee Owned Features and Components Digital Surface Resistance Meter Figure 3 Statguard Flooring D19290 Digital Surface Resistance Meter Figure 2 Statguard Flooring D19290 Digital Surface Resistance Meter Kit Figure 4 Digital Surface Resistance Meter features and components C B D E E G F A ...

Page 3: ...ings and Memory Recall menus to exit and return to the home screen G Battery Compartment Open this compartment to install the four AA alkaline batteries needed to power the meter Replace the batteries once the battery icon on the display is empty Home Test Results Screen H J L I K Figure 5 Home Test Results screen H Battery Life Indicator Displays the approximate life of the meter s 4 AA alkaline ...

Page 4: ...ctrode placements include Most commonly used area of a surface Most worn area Center of surface Furthest area from a grounded point If the surface to be measured has sections i e floor tiles garment panels place the 5 pound electrodes on different sections for Rtt measurements Clean the material s surface for test lab measurements but do not clean the surface for materials that are already install...

Page 5: ...5 pound electrodes and place the electrodes on the most commonly used area of the surface Use the 10 electrode spacer to space them 10 apart from each other The electrodes should also be 2 away from any edge and 3 away from any grounded point If the most used area is not obvious use two points near the center of the surface 4 Push the red pushbutton to perform a measurement Should the 15 second el...

Page 6: ...n cleaning instructions Per ANSI ESD S4 1 Worksurfaces The test specimens and electrodes shall be cleaned twice with a minimum 70 isopropanol water solution using a clean low linting cloth each time Note The item should then be conditioned for 72 hours minimum Figure 13 Measuring the Resistance Point to Point Rtt of floor materials Specifications Resistance Ranges 1 x 103 to 1 x 106 ohms 10 Volts ...

Reviews: