Calibration

Frequency of recalibration should be based on the 

critical nature of those ESD sensitive items handled 

and the risk of failure for the ESD protective equipment 

and materials. In general, Desco recommends that 

calibration be performed annually.
Use the Desco 

98221

 Wave Distortion Monitor 

Verification Tester to perform periodic verification (once 

every 6-12 months) of the Mini Monitor. The verification 

tester can be used to check the test limits of the Mini 

Monitor without removing it from the workstation.
See 

TB-3074

 for more information.

Operation

1.  Fit the wristband snugly onto the wrist. 
2.  Snap the wrist cord to the wristband. 
3.  Plug the wrist cord into the operator jack. The green 

operator LED will illuminate. This indicates that the 

operator is properly grounded. 

4.  If this does not happen, examine the wrist cord for 

continuity or damage and the wristband to ensure 

that it fits securely. If you have dry skin, apply an 

approved dissipative hand lotion such as 

Menda 

Reztore® ESD Hand Lotion

5.  When leaving the workstation, first touch the 

monitor's park snap. The monitor's audible alarm 

will disable for approximately 10 seconds. Use this 

time to disconnect the wrist cord from the wristband 

and attach it to the park snap without sounding a 

false alarm. 

6.  Upon returning to the workstation and disconnecting 

the wrist cord from the park snap, the monitor will 

provide approximately 10 seconds for the operator 

to reconnect the wrist cord to a worn wristband 

before sounding the monitor's alarm.

Figure 9. Desco 

98221

 Wave Distortion Monitor 

Verification Tester

Figure 8. Using the Mini Monitor

Summary of Contents for 19258

Page 1: ...he continuous or constant monitor The test voltage creates a sine wave that the monitor circuit compares to established patterns By monitoring the distortions or shape of the sine wave Wave Distortion Technology determines if the monitored circuit is complete the operator is in the circuit and the total equivalent DC resistance is within specifications Wave Distortion Technology produces a very fa...

Page 2: ...ashers 2 Flat Head Screw 6 32 x 1 4 1 Power Adapter 24VDC North America Plug 1 Certificate of Calibration 19243 Mini Monitor Universal Power Adapter 1 Mini Monitor 2 Countersunk Washers 2 Flat Head Screw 6 32 x 1 4 1 10mm Interchangeable Park Snap 1 10mm Operator Jack Adapter 1 Power Adapter 24VDC IEC Inlet 1 Certificate of Calibration Figure 2 North American power adapter included with the 19239 ...

Page 3: ...park a wrist strap when leaving the workstation without creating a false alarm Touch the park snap and the monitor s audible alarm will disable for approximately 10 seconds Use this time to disconnect the wrist cord from the wristband and attach it to the park snap or remove the banana plug from the jack Upon returning to the workstation and disconnecting the wrist cord from the park snap the oper...

Page 4: ...ark snap and 10mm banana jack adapter for operators who use wrist cords with 10mm terminations Use the park snaps knurled rims to unscrew the 4mm park snap from the monitor and install the 10mm park snap to the monitor Plug the 10mm operator jack adapter into the monitor s operator jack Installation 1 Remove the Mini Monitor from the carton and inspect for damage 2 Locate an appropriate AC outlet ...

Page 5: ... 2 Snap the wrist cord to the wristband 3 Plug the wrist cord into the operator jack The green operator LED will illuminate This indicates that the operator is properly grounded 4 If this does not happen examine the wrist cord for continuity or damage and the wristband to ensure that it fits securely If you have dry skin apply an approved dissipative hand lotion such as Menda Reztore ESD Hand Loti...

Page 6: ... megohms 20 Operator Test Voltage 1 2V peak to peak 1 2µA open circuit Worksurface Test Voltage 5 to 7 5V open circuit Alarm Response Time 50 ms Country of Origin United States of America This cannot be verified with standard DC test equipment The continuous monitor is an impedance sensing device and the limits are determined by the magnitude and angle of the impedance The worksurface mat must hav...