16
A q u a - B o x O w n e r ’ s M a n u a l
www.deroyal.com
800.251.9864
RIGID CANISTERS
1. Transport full canisters by cart or roll-stand to the Aqua-
Box using PPE’s in accordance with hospital policy.
2. Ensure that the appropriate Evacuation Tube is attached
to the end of the suction hose (C).
3. If fluid treatment option is used, verify that there is
sodium hypochlorite (bleach) and that the yellow tube (F)
is submerged in the bleach. Check to see if the valve (E) for
bleach line is open (vertically oriented).
4. Open pour spout of the suction canister(s).
5. Insert the Evacuation Tube and seal the port with the
purple cap.
6. Turn the unit “On”.
a.
For Models 1500 and 1300, press the “Start” button
(A). If the Aqua-Box shuts off while evacuating,
simply press “Start” button (A) again.
b. For Models 1200 and 1100, turn the valve on the
right side of the unit to the “On” position.
7. Open Thumb valve (D) to begin evacuation.
INSTRUCTIONS FOR USE
NON-RINSING UNITS
A - “Start” Switch
B - “Kill” Switch
C - Suction Hose (Grey)
D - Thumb Valve
E - Treatment Solution Valve
F - Treatment Solution Tube (Yellow)
G - Suction Hose Hanger
A
B
D
G
C
E
F
Warning: As whenever handling any infections liquid waste, PPE’s should be worn at all times in accordance with facility policies.