Tare Memory/Net Weighing
Tare Memory/Net Weighing (Menu Code 216)
Symbol displayed: NET
Use this program to determine the individual weight of several samples
and record the net weight.
1. Select menu code 2 1 6
2. Press the FUNCTION key to store the tare weight
3. Add the next sample, press the FUNCTION key to store the weight and
tare the balance
4. Repeat steps 2 - 3, as needed
5. When completed, press and hold CF key for at least 2 seconds
Output Parameter
- print individual components (N1) & tare weights
(Menu Code 7 3 1*)
- print net total tare weight (T1)
(Menu Code 7 3 2)
Output Symbols
T1
tare weight stored in memory
N1
net weight when tare weight is stored
N
gross weight = tare + net weight
Balance Operating Parameters
In the operating menu of the balance, you can define how your balance
will adapt to ambient conditions and also how it will work to meet your
special requirements.
Adapting the Balance to Ambient Conditions
The balance can be adapted to the prevailing ambient conditions at the
place of installation:
- very stable conditions
(Menu Code 1 1 1)
- stable conditions
(Menu Code 1 1 2)
- unstable conditions
(Menu Code 1 1 3)
- very unstable conditions
(Menu Code 1 1 4)
Standard Weighing Mode – Manual Filling Mode
You can optimally adapt your balance to meet either of these
requirements.
In the manual filling mode, the display compensates for fluctuations of the
load on the balance, giving you especially fast and stable readouts.
- standard weighing mode
(Menu Code 1 2 1*)
- manual filling mode
(Menu Code 1 2 2)
12