Section 10:  Troubleshooting

Although service and repair of the Cavitron

 JET Combination

System should be performed by DENTSPLY

®

 personnel, the

following are some basic trouble shooting procedures that will
help avoid unnecessary service calls. Generally, check all lines and
connections to and from the System, a loose plug or connection
will often create problems. Check the settings on the System’s
knobs.

Section 9: System Care cont'd

• Turn System OFF.
• Remove the Powder Chamber.
• Remove filter from holder and remove the tubing from

one end of the filter.

• Disassemble the filter as shown.
• Inspect the filter element. Replace if necessary.
• Reassemble the filter and reconnect the tubing and replace

filter in holder.

• Inspect second filter in a similar manner.
• Replace the Powder Chamber.
• Turn the System ON.

NOTE:

 Never operate the System without all filters in place.

10.1 Troubleshooting Guide

System will not operate:

(Power Indicator Light is not lighted)

1. Check that the Power switch is in the ON position, and that

detachable Power Cord is fully seated in the receptacle on back
of System.

2. Check that the System’s three-prong plug is fully seated in an

appropriate AC receptacle, and that AC current is present.

(Power Indicator Light is lighted)

1. Check that the Foot Control Connector is fully seated in the

Foot Control Receptacle on the back of the System.

System operates:

(No lavage flow at insert tip)
1. Assure that Lavage control is properly adjusted.
2. Check that irrigant supply control valve/s (dental office water

supply) are open.

3. If connected to DualSelect™ Dispensing System, check that

fluid level  in the selected bottle is sufficient. Make sure valves
are open when using external water source.

(JET Insert Nozzle  blocks repeatedly)
1. Powder is moisture contaminated (lumpy). Discard powder.
2. Air filter is contaminated.  Clean or replace filter.

(No bleed air)
1. Blocked JET nozzle.  Clean JET Nozzle using supplied tool.
2. Blocked filter.  Clean or replace filter.

(No or poor cleaning action)
1. Very low powder level or empty Powder chamber.  Fill Powder

Chamber.

2. Blocked JET Nozzle.  Clean JET Nozzle using supplied tool.
3. Fill Cap is loose. Tighten powder fill cap, and turn System

ON.

(Continuous powder agitation)

Fill cap not securely sealed. Shut System OFF.  Remove the
O-ring seal from the fill cap and clean residual powder from the
cap. Be careful not to scratch or otherwise damage the plastic cap.
Wipe off the O-ring and place it in the fill cap and turn System
ON.

10.2 Technical Support and Repairs

Pointer

View Window

Fill Cap

Dump/Pressure

Line

Air-Powder

Flow

Air Inlet

Air Supply Nozzle

Filter

(depressurizing System)

Powder Bowl

For technical support and assistance call 1-800-347-7412
Monday through Friday, 8:00 AM to 5:00 PM (Eastern Time).
For other areas, contact your local DENTSPLY

®

 representative.

Summary of Contents for Cavitron JET

Page 1: ...oigneusement et compl tement avant d utiliser le syst me Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad Syst me combin Cavitron JET D tartreur ultras...

Page 2: ...e insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional oper...

Page 3: ...nection 6 Cavitron JET Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System 6 8 6 1 System Controls 6 2 Handpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishin...

Page 4: ...Do not direct the air polishing stream at soft tissue or into the sulcus Tissue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulc...

Page 5: ...ystem s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patien...

Page 6: ...Specifications 5 7 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated 5 5 Unpacking the System Carefull...

Page 7: ...upply line to office air supply line or a Cavitron DualSelect dispensing System Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaleran...

Page 8: ...e automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs air powder flow to insert tip Insert Heater Rod Heats the delivered water for patient comfort O Ring...

Page 9: ...perationforUltrasonicProphylaxisProcedures No ultrasonic energy or lavage flow Ultrasonics and lavage flow are activated Foot control depressed all the way 6 7 Operation for Air Polishing Procedures F...

Page 10: ...he rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE Set the Unit s Power Adjustment knob to the lowest power s...

Page 11: ...ing of the powder chamber An additional cap without a hole is provided for storage It is strongly recommended that the Powder Chamber be emptied at the end of each day This will reduce moisture absorp...

Page 12: ...ly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes re...

Page 13: ...t s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Use a saliva ejector or HVE In the posterior regions use the cheek to help contain aerosol spray 8 7 Perf...

Page 14: ...ectant solution to air dry 5 If using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut Down Procedures at the end of the day 1 Turn the Unit OFF 2 Rem...

Page 15: ...in the Foot Control Receptacle on the back of the System System operates No lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s de...

Page 16: ...Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for ope...

Page 17: ...6 Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Ph...

Reviews: