background image

ENGLISH  • 9

Section 7:  Accessories

Section 8:  Techniques for Use

8.1 Patient Positioning

• For optimal access to both the upper and lower arches,

the backrest of the chair should be adjusted as for other
dental procedures. This assures patient comfort and
clinician visibility.

7.1 Accessories List*

1. AC Power Cord Set
2. Cavitron

®

 Ultrasonic Inserts

3. Cavitron™ Jet Insert Nozzle
4. Dual position Foot Switch  Assembly
5. DualSelect

 Medicament Dispensing System

6. JetShield™ Aerosol Reduction Device

*For detailed information, contact your local DENTSPLY

®

 Representative.

• Have the patient turn their head to the right or left.  Also

position their chin up or down  depending upon the
quadrant and surface being treated. Evacuate irrigant
using either a saliva ejector or HVE.

8.2 Performing Ultrasonic Scaling

Procedures

NOTE: Refer to the Infection Control Information card (Form
no. 80722) supplied with your Unit for general procedures to be
followed at the beginning of each day and between patients.

• The edges of Cavitron

®

 Ultrasonic Inserts are intention-

ally rounded so there is little danger of tissue laceration
with proper ultrasonic scaling technique. Whenever the
insert tip is placed within the patient’s mouth, the lips,
cheek and tongue should be retracted to prevent acciden-
tal prolonged contact with the activated tip.

• Hold the empty Handpiece in an upright position and

activate the Foot Control until fluid exits.

• Lubricate the rubber O-ring on the insert with water

before placing it into the handpiece.  Fully seat insert with
a gentle push-twist motion.  DO NOT FORCE.

• Set the Unit’s Power Adjustment knob to the lowest

power setting for the application and the selected insert.

• Activate the  Foot Control and hold the Handpiece over a

sink or drain. Check spray temperature to verify fluid is
reaching the working end of the insert tip. Adjust the
Lavage flow knob to ensure adequate flow for the
selected Power setting. Greater flow settings provide
cooler irrigant.

• It may be necessary to adjust lavage with the System in

“Boost” mode (Foot Control fully depressed) so
adequate fluid will be available to cool tip, tooth interface.

• In general, it is suggested that a “feather-light-touch” be

used both supra- and subgingivally. The motion of the
activated tip and acoustic effects in the irrigating fluid, in
most cases, is adequate to remove even the most
tenacious calculus.

• Use Cavitron

®

 efficiency indicator if an insert seems to

require increased finger pressure to function adequately.

• The use of a saliva ejector or HVE is recommended

during all procedures.

8.3 Patient Comfort Considerations

Reasons for sensitivity

1. Incorrect tip placement. Point should be directed away

from root surfaces.

2. Not keeping tip in motion on tooth. Do not allow the

insert to remain in a static position on any one area of
the tooth. Change the insert’s path of motion.

3. Applying pressure. Use extremely light grasp and

pressure, especially on exposed cementum.

4. If sensitivity persists, decrease power setting and/or

move from the sensitive tooth to another and then
return.

Summary of Contents for Cavitron JET

Page 1: ...oigneusement et compl tement avant d utiliser le syst me Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad Syst me combin Cavitron JET D tartreur ultras...

Page 2: ...e insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional oper...

Page 3: ...nection 6 Cavitron JET Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System 6 8 6 1 System Controls 6 2 Handpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishin...

Page 4: ...Do not direct the air polishing stream at soft tissue or into the sulcus Tissue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulc...

Page 5: ...ystem s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patien...

Page 6: ...Specifications 5 7 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated 5 5 Unpacking the System Carefull...

Page 7: ...upply line to office air supply line or a Cavitron DualSelect dispensing System Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaleran...

Page 8: ...e automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs air powder flow to insert tip Insert Heater Rod Heats the delivered water for patient comfort O Ring...

Page 9: ...perationforUltrasonicProphylaxisProcedures No ultrasonic energy or lavage flow Ultrasonics and lavage flow are activated Foot control depressed all the way 6 7 Operation for Air Polishing Procedures F...

Page 10: ...he rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE Set the Unit s Power Adjustment knob to the lowest power s...

Page 11: ...ing of the powder chamber An additional cap without a hole is provided for storage It is strongly recommended that the Powder Chamber be emptied at the end of each day This will reduce moisture absorp...

Page 12: ...ly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes re...

Page 13: ...t s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Use a saliva ejector or HVE In the posterior regions use the cheek to help contain aerosol spray 8 7 Perf...

Page 14: ...ectant solution to air dry 5 If using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut Down Procedures at the end of the day 1 Turn the Unit OFF 2 Rem...

Page 15: ...in the Foot Control Receptacle on the back of the System System operates No lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s de...

Page 16: ...Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for ope...

Page 17: ...6 Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Ph...

Reviews: