Section 9:  System Care

9.1 Daily Maintenance

Start-Up Procedures at the beginning of the day:
1. Open the manual shut-off valve on the dental water supply

system.

2. With the system OFF, unscrew the powder chamber cap.

Verify the powder bowl is empty.  Turn the Unit ON for 15
seconds to eliminate residual powder or moisture in the lines.
Turn the system OFF.

3. Shake the powder in the storage jar to create an even consis-

tency of powder mixture.

4. Pour the powder into the chamber until it reaches the top of

the center tube.

5. Secure the lid on the powder chamber.
6. Turn the system ON. Hold the Handpiece (without an insert

installed) over a sink or drain, activate the Foot Control, and
flush the water line at maximum water flow for at least 2
minutes.

7. Place a sterilized Cavitron

®

 Ultrasonic Insert or Cavitron™

JET Insert Nozzle into the Handpiece and adjust the Power
and Lavage Flow as required.  For the JET mode, the powder
flow can be adjusted.

Between Patients:
1. Remove Cavitron Ultrasonic Insert or Cavitron™ JET Insert

Nozzle (clean using the wire cleaning tool).  Sterilize.

2. Set Water Control to highest setting.   Activate the System and

flush at least 30 seconds.

3. Turn the System OFF.
4. Clean and disinfect the surfaces of the cabinet, Power Cord,

Handpiece Connector and cable assembly, Foot Control and
cable assembly by applying an approved non-immersion type
disinfectant solution* carefully following the instructions
provided by the disinfectant solution manufacturer. To clean
system, generously spray disinfectant solution on a clean towel
and wipe all surfaces. Discard used towel. To disinfect system,
generously spray disinfectant on a clean towel and wipe all
surfaces.  Allow disinfectant solution to air dry.

5. If using a closed water supply, check for adequate fluid

volumes for next patient’s scheduled treatment.

Shut-Down Procedures at the end of the day:
1. Turn the Unit OFF.
2. Remove Cavitron

®

 Ultrasonic Insert or Cavitron™ JET Insert

Nozzle (clean using the wire cleaning tool).  Sterilize.

3. Unscrew the powder chamber cap.
4. Remove the powder chamber from the cabinet.  Pour the

powder into the powder storage jar and secure the lid.

5. Hold the open end of the powder chamber away from you.

Activate  the System for 15 seconds to clear the powder
chamber.  A high volume evacuator can be used to remove any
residual powder.

6. Secure the lid on the powder chamber.
7. Clean and disinfect the surfaces of the cabinet, Power Cord,

Handpiece Connector and cable assembly, Foot Control and
cable assembly by applying an approved non-immersion type
disinfectant solution* carefully following the instructions

9.2 Periodic Maintenance

• The two repairable filters should be disassembled and

 inspected every 90 days.

"O" Ring

Duckbill

Filter

Repairable Filters:

provided by the disinfectant solution manufacturer. To clean
system, generously spray disinfectant solution on a clean towel
and wipe all surfaces. Discard used towel. To disinfect system,
generously spray disinfectant on a clean towel and wipe all
surfaces.  Allow disinfectant solution to air dry.

*

NOTE:

 Water-based disinfection solutions are preferred. Some

alcohol-based disinfectant solutions may be harmful and may
discolor plastic materials.

If the System is connected to a DualSelect™ Medicament
Dispensing System follow the recommended procedures listed in
the DualSelect™ manual.

If powder fluffing is observed when Foot Control is not in use,
this would indicate an air leak. To correct, turn the Unit OFF,
remove the Fill Cap, clean any residual powder from the O-ring
seal and threads, replace the Fill Cap, tighten, and turn the Unit
back ON.

Summary of Contents for Cavitron JET

Page 1: ...oigneusement et compl tement avant d utiliser le syst me Por favor lea la totalidad de las instrucciones cuidadosamente antes de hacer funcionar la unidad Syst me combin Cavitron JET D tartreur ultras...

Page 2: ...e insert s working tip This combined with acoustic effects of the coolant water produces a synergistic action that literally powers away the heaviest calculus deposits while providing exceptional oper...

Page 3: ...nection 6 Cavitron JET Combination System Ultrasonic Scaler and Air Polishing Prophylaxis System 6 8 6 1 System Controls 6 2 Handpiece 6 3 Cavitron 30K Ultrasonic Inserts 6 4 Cavitron JET Air Polishin...

Page 4: ...Do not direct the air polishing stream at soft tissue or into the sulcus Tissue emphysema has been reportedly caused when the air water powder stream was directed at the soft tissues or into the sulc...

Page 5: ...ystem s aerosol dispersion can be effectively controlled and minimized Please carefully follow the procedural guidelines in this manual regarding the use of your System For maximal operator and patien...

Page 6: ...Specifications 5 7 Foot Control Cable Assembly Connection Align the Foot Control plug with the receptacle on the back of the System and push it in until firmly seated 5 5 Unpacking the System Carefull...

Page 7: ...upply line to office air supply line or a Cavitron DualSelect dispensing System Inspect all connections to make certain there are no leaks Section 6 Cavitron Jet Combination System Ultrasonic Scaleran...

Page 8: ...e automatically selects the JET mode when placed in handpiece Cleaning Powder Delivery Tube Directs air powder flow to insert tip Insert Heater Rod Heats the delivered water for patient comfort O Ring...

Page 9: ...perationforUltrasonicProphylaxisProcedures No ultrasonic energy or lavage flow Ultrasonics and lavage flow are activated Foot control depressed all the way 6 7 Operation for Air Polishing Procedures F...

Page 10: ...he rubber O ring on the insert with water before placing it into the handpiece Fully seat insert with a gentle push twist motion DO NOT FORCE Set the Unit s Power Adjustment knob to the lowest power s...

Page 11: ...ing of the powder chamber An additional cap without a hole is provided for storage It is strongly recommended that the Powder Chamber be emptied at the end of each day This will reduce moisture absorp...

Page 12: ...ly Pivot nozzle inside the cup to adapt to incisal gingival mesial and distal surfaces Apply two seconds of spray for each segment of the tooth releasing the Foot Control each time If spray escapes re...

Page 13: ...t s head toward you to help prevent puddling in the cupped lip and minimize aerosol dispersion Use a saliva ejector or HVE In the posterior regions use the cheek to help contain aerosol spray 8 7 Perf...

Page 14: ...ectant solution to air dry 5 If using a closed water supply check for adequate fluid volumes for next patient s scheduled treatment Shut Down Procedures at the end of the day 1 Turn the Unit OFF 2 Rem...

Page 15: ...in the Foot Control Receptacle on the back of the System System operates No lavage flow at insert tip 1 Assure that Lavage control is properly adjusted 2 Check that irrigant supply control valve s de...

Page 16: ...Weight 5 4 kg Dimensions Height 10 2 cm Width 18 7 cm Depth 24 1 cm Handpiece Cable 2 0 m Foot Control Cable 2 4 m Power Cord 2 1 m Water Supply Line 2 4 m Air Supply Line 2 5 m Footswitch Not for ope...

Page 17: ...6 Australia DENTSPLY Australia Pty Ltd 204 206 Gipps Street Abbotsford Victoria 3067 Australia Phone 3 9417 1666 3 9417 1777 Canada DENTSPLY Canada 161 Vinyl Court Woodbridge Ontario L4L 4A3 Canada Ph...

Reviews: