16
Problem:
Console does not light when
turned on:
Console lights when turned on,
but handpiece does not turn:
Improper display:
Motor slowing down or sluggish:
Correction:
If lights are blinking, press a button or foot switch
to exit Sleep Mode.
Check console to power connection.
Check voltage setting.
Check fuse. If blown, replace with 1.0A/250V
slo-blo fuse for 115 volts, & 0.5A/250V slo-blo
fuse for 230 volts.
Check motor plug connection.
Check foot switch connection.
Depress foot switch.
Check that a file is properly seated in the
handpiece and the latch is closed.
Verify that ratio setting matches handpiece ratio.
Use Calibration function.
Turn power switch off, wait 5 seconds, then turn
back on to reset.
Dirty, under-lubricated handpiece.
Handpiece lubricant is running into motor.
After lubricating and before autoclaving, set
handpiece with head down to let excess lubricant
drain out.
Check voltage setting.
TROUBLESHOOTING: