13

7. The display will show:

ACCESS FILE LIBRARY?
1=yes, 2=no, 3=esc

Press Button #2.  Do not access
the file library.

8. The display will show:

SAVE CURRENT SETTING
1=yes, 2=no, 3=esc

Press Button #1 to record settings on the
Preset Button.

9. The display will resemble the following:

(The handpiece ratio, speed and torque
values will vary according to your settings).

USER DEFINED PRESET
1/8 AG     300     240

The Preset Button is now programmed with
the specified settings.

10. When you have completed the program-

ming of the desired preset buttons,
re-enable the Preset Memory Lock to avoid
inadvertently entering the File Library
during a procedure.  For instructions, read
about the System Setup Menu beginning
on page 6.

1. Disable the Preset Memory Lock and

enable or disable Auto Stop Reverse and
torque warning tone.
For instructions, read about the System
Setup Menu beginning on page 6.

2. Set the Handpiece Ratio to the correct set-

ting.
Press the Handpiece Ratio Selector up or
down arrows until the Handpiece Ratio
Indicator matches the ratio of the
handpiece to be used.

3. Set the speed to the desired setting.

Press the Speed Selector up/down arrows
until the desired speed is shown on the
display.

4. Set the torque level to the desired setting.

Press the Torque Selector up/down arrows
until the desired torque is shown on the
display.
To operate the motor at full torque, press
the "MAX" Torque Button.  The LED above
the "MAX" Torque Button will display a
steady red light, the Torque Indicator on the
display will read "MAX".  Operating at max.
torque will override the Auto Stop Reverse
feature.

5. Set Handpiece Rotation to the desired

direction.
Press the Forward/Reverse Selector to
change the direction of handpiece rotation.
A red L.E.D. light above the
Forward/Reverse Selector indicates
reverse rotation.

6. Press and hold the desired Preset Button

for 1 second (until you hear a double
beep).

11. PROGRAMMING A PRESET BUTTON WITH A USER DEFINED PRESET

Summary of Contents for AEU-25T

Page 1: ...Endo DTC DIGITAL TORQUE CONTROL DUAL VOLTAGE AEU 25T Electronic Endodontic System OPERATION and MAINTENANCE INSTRUCTION MANUAL...

Page 2: ...electric handpiece control system that is used in dentistry to drive endodontic files and instruments CONTRAINDICATIONS None Known WARNINGS Equipment not suitable for use in the presence of a flammabl...

Page 3: ...se features combine to make the perfect rotary file system for endodontic and other dental procedures Congratulations This system is engineered to provide many years of reliable service Please read th...

Page 4: ...e AE 4B 30 motor to the receptacle on the lower front panel of the console 4 Attach the E Type handpiece to the motor See Figure 2 5 Attach the supplied AE 7P On Off Foot Switch to the connector on th...

Page 5: ...e Front Figure 4 Console Back Handpiece Ratio Selector Speed Selector Calibration System Setup PRESET MEMORY LOCK RECALL FACTORY SETUP AUTO STOP REVERSE REVERSE BEEPER TORQUE WARNING TONE ACCESSORY WA...

Page 6: ...an endodontic rotary file into the handpiece 4 Use the Calibration function to calibrate the Endo DTC See Caution at right Press and release the CAL button The unit will run for a few seconds measuri...

Page 7: ...sal S2 F1 File 3 ProTaper Universal F2 F3 F4 F5 File 4 ProTaper Universal Retreatment D1 D2 File 5 ProTaper Universal Retreatment D3 Preset Button 4 Default File Series DENTSPLY Tulsa Dental Specialti...

Page 8: ...strument breakage is greatly reduced Torque settings are held in memory even after unit is turned off For most endodontic root canal applications the included 1 8 reduction contra angle handpiece is n...

Page 9: ...To exit the System Setup Menu press 3 SAVE CHANGES selection will appear Select 1 to save changes and exit or select 2 to reject all changes and exit 6 CALIBRATION AND SYSTEM SETUP BUTTON Calibration...

Page 10: ...dible when a GT Accessory file or ITR Emulation Accessory file is in use This is a system wide setting and is not saved on a Preset Button when programming To disable the Accessory Warning press 1 To...

Page 11: ...e file series shown in Chart 1 on page 5 The Preset Buttons are also used to navigate the various menus in the System Setup and File Library The Preset Buttons can be customized to suit individual pre...

Page 12: ...you have completed the customiza tion of the desired preset buttons re enable the Preset Memory Lock to avoid inadvertently entering the File Library during a procedure For instructions read about the...

Page 13: ......

Page 14: ...d inadvertently entering the File Library dur ing a procedure For instructions read about the System Setup Menu beginning on page 6 1 Disable the Preset Memory Lock For instructions read about the Sys...

Page 15: ...ge 6 2 Set the Handpiece Ratio to the correct set ting Press the Handpiece Ratio Selector up or down arrows until the Handpiece Ratio Indicator matches the ratio of the handpiece to be used 3 Set the...

Page 16: ...e from the motor and install plug as shown at right The included motor handpiece holding cradle is also fully autoclavable CONSOLE The exterior of the console may be cleaned by wiping with a soft clot...

Page 17: ...voltage replace the fuses with fuses that are rated for the new voltage Turn the Fuse holder until the white lettering with the proper voltage is right side up 4 Replace the Fuse holder and close the...

Page 18: ...5 volts 0 5A 250V slo blo fuse for 230 volts Check motor plug connection Check foot switch connection Depress foot switch Check that a file is properly seated in the handpiece and the latch is closed...

Page 19: ...anty any risks or liabilities arising from the clinical use of its products whether or not such use involves coincidental utilization of products manufactured by others Tulsa Dental Specialties makes...

Page 20: ...Johnson City TN 37604 1 800 662 1202 Fax 1 800 597 2779 www TulsaDentalSpecialties com Rx Only DFUDTC Rev 4 12 12 P N 420322 Rev J ECO 12954 12 12 Software P N 890002 Rev J TUL06 Series 2012 DENTSPLY...

Reviews: