![Dentsply Sirona LEDview Plus Operating Instructions Manual Download Page 36](http://html1.mh-extra.com/html/dentsply-sirona/ledview-plus/ledview-plus_operating-instructions-manual_2483670036.webp)
65 67 510 D3642
36
D3642
.
201.01.03
.
02 01.2018
4 Operation
Dentsply Sirona
4.5 Accessories for the operating light
Operating Instructions LEDview Plus
4.5.2
Patient mirror
General information on the patient mirror
The accessory interface must be installed on the operating light in order
to use the patient mirror.
You can install the patient mirror on the mounting adapter of the
accessory interface and remove it again when not in use.
Mounting the patient mirror
✔
The accessory interface is installed on the operating light.
1. Set the bayonet mount of the patient mirror at a 45° angle at the
mounting adapter.
2. Turn the patient mirror clockwise until it locks into place.
3. Check that the patient mirror is securely positioned.
CAUTION
The operating light can be accidentally switched off via the contactless
sensor and the patient dazzled when using the mirror in the proximity of
the mouth.
Before using the mirror in the proximity of the mouth, switch the
operating light off and switch off the contactless sensor function via the
treatment center keyboard or via the integrated keyboard (see sections
"Operation via dentist or assistant element" [ → 22] or "Operation via
integrated keyboard" [ → 31]).