![Dentsply Sirona LEDview Plus Operating Instructions Manual Download Page 15](http://html1.mh-extra.com/html/dentsply-sirona/ledview-plus/ledview-plus_operating-instructions-manual_2483670015.webp)
65 67 510 D3642
D3642
.
201.01.03
.
02 01.2018
15
Dentsply Sirona
3 Unit description
Operating Instructions LEDview Plus
3.2 Technical data
båÖäáëÜ
Light field size:
Approx. 200 mm x 100 mm at distance of
700 mm, measured from the light emission
area of the light
Light distribution
The following image shows the light
distribution at 40000 lux.
Area A: min. 75% illuminance according to ISO 9680 standard
Area B: min. 50% illuminance according to ISO 9680 standard
75% / 50% / 10%: Measured illuminance with the LEDview Plus
operating light
Protection class:
Unit model: The unit supplying power to the
LEDview Plus corresponds to protection
class I
Ceiling model: Class I equipment
Device class in
accordance with Directive
93/42/EEC:
LEDview Plus: Class I equipment
Ceiling model with LEDview Plus: Class I
equipment
Degree of protection
against ingress of water:
Ordinary equipment (IP X0 – without
protection against ingress of water)
Operating mode:
Continuous operation
Transport and storage
conditions:
Temperature: -40 °C – +70 °C
(-40 °F – +158 °F)
Relative humidity: 10 % - 95 %
Air pressure: 50 kPa – 106 kPa
Operating conditions:
Ambient temperature: 10°C – 35°C
(50 °F – 95 °F)
Relative humidity: 30 % - 85 %
no condensation
Air pressure: 70 kPa – 106 kPa
Installation location: ≤ 3,000 m above sea
level