![Dentsply Sirona inEos X5 Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/dentsply-sirona/ineos-x5/ineos-x5_operating-instructions-manual_2483661033.webp)
Dentsply Sirona
Operating Instructions inEos X5
7 Controls and displays
7.2 Operating state LED
64 17 096 D3586
D3586.201.01.11.02 07.2018
33
7.2
Operating state LED
The LED shows the operating status of the unit:
NOTE
Once the unit is switched on, the LED must light up in one of the
following colors. Should this not be the case, this indicates a unit
defect, which must be eliminated prior to further use.
LED lights up green
The unit is ready for operation.
The articulated arm will move only in case of user interaction in the
software or if the Start key is actuated.
LED lights up yellow
An automatic acquisition process has been started; the arm is already
moving or is about to start moving.
CAUTION
Risk of injury!
Do not place any part of your body either wholly or partially in the
working area of the articulated arm in this operating state!
Ensure that there are no objects and/or body parts on the work plate
(shaded area) during automatic operation. Otherwise this creates a
collision hazard as a result of the swiveling movement of the
articulated arm.
LED lights up red
Error state.
Causes may include:
● Activation of safety light barrier on the scan element
● Pressing and holding the Start key
● Malfunctioning of drives
The articulated arm will not move in this operating state.
Proceed as prompted by the inLab software.