4 Transport and commissioning
Dentsply Sirona
4.6 Commissioning
Operating Instructions CEREC SpeedFire
4.6.4.3
Manual unit search
✔
The unit must be switched on.
✔
The unit must be directly connected to the existing network with the
network cable.
1. Start the CEREC SW software.
2. Move the mouse cursor to the top of the window.
or
➢
Click the start window button.
ª
The system menu is displayed.
3. Click on the
"Configuration" button in the system menu.
4. Click on the
"Devices" button.
5. Click on the
"Add Device (Manual)" button.
6. Select
"Network" .
7. Enter the network address.
8. Click on the
"Ok" button.
ª
The software attempts to contact the device.
9. Enter a name for the new unit.
If connection fails, check the network status.
4.6.4.4
Remove the unit
✔
The unit is not in operation.
1. Click on the
"Configuration" button in the system menu.
2. Click on the
"Devices" button.
3. Click on the unit that you wish to uninstall.
4. Click on the
"Delete Device" button.
ª
You will be asked if you would like to remove the unit.
5. Click on the
"YES" button.
ª
The device is removed.
4.6.5
Configuration following initial startup of the unit
Here you will be given the option to select the required language and
temperature unit and to set the clock.
The configuration can be found in the
"Settings" /"basic settings" menu.