FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
- 37 / 208 -
Error
code
Error
Cause
Check/Remedy
During
rotation of
the motor
handpiece
E-00 Self-Check
Malfunction of circuit.
Contact your distributor.
E-01 Over current
The motor handpiece is locked
(at the time of the auto reverse
mode).
Remove load.
The motor handpiece cord has
shorted.
Contact your distributor.
E-02 Over voltage
Malfunction of circuit.
Contact your distributor.
E-03 Motor handpiece
sensor
The motor handpiece cord is
disconnected.
Connect the motor
handpiece cord correctly.
Faulty sensor (Hall IC).
Severed cord (signal line).
Contact your distributor.
E-04 Overheating of
motor
High load was continuously
applied to the motor handpiece
for a relatively long period of
time.
Let the motor cool down
before resuming
operation.
E-05 PAM circuit
Abnormal voltage generated in
start / stop circuit.
Faulty start / stop circuit from
PAM (L Slide).
Contact your distributor.
E-06 Rotor lock
The motor handpiece is locked
at the time of startup.
Remove load.
Faulty contra-angle.
Motor faulty.
Faulty sensor (Hall IC).
Severed cord (Signal, Power
line).
Contact your distributor.
E-08 Over current
Shorted cord (power line).
Shortstop of the motor winding.
Contact your distributor.
E-09 ITRIP
Faulty motor and circuit.
Contact your distributor.
At the time
of charging
E-10 Battery current
The battery current is too low
or too high. The battery is flat
or not inserted.
Put the battery into the
battery compartment or
replace the battery.
E-11 Display
Faulty display driver.
Contact your distributor.
E-12 Low battery voltage The battery voltage is too low.
The battery is not inserted or
the battery operating life is
over.
Put the battery into the
battery compartment or
replace the battery.
E-13 High battery
voltage
The battery voltage is too high
(circuit malfunction).
Contact your distributor.
E-14 Beyond the range
of working
temperature
Beyond the range of working
temperature or damage to the
thermistor in the battery
section.
Use within the range of
working temperature or
replace the battery.
Summary of Contents for X-SMART Plus
Page 1: ...USER MANUAL...
Page 3: ...FISDR JAP F1902124 X 00 2011 created 07 2011 4 208...
Page 83: ...FISDR JAP F1902124 X 00 2011 created 07 2011 84 208...
Page 163: ...FISDR JAP F1902124 X 00 2011 created 07 2011 164 208...
Page 207: ...FISDR JAP F1902124 X 00 2011 created 07 2011 208 208...
Page 208: ...FISDR JAP F1902124 X 00 2011 created 07 2011 209 208...