FISDR-JAP / F1902124.X / 00 / 2011 – created 07/2011
- 14 / 208 -
9
AUTO REV
Only for continuous rotary systems, selects one of the 3 auto
reverse modes (see chapter 6.6.4. Auto Reverse Function):
AUTO REVERSING: Automatic stop and reverse motion followed
by forward rotation when preset torque is reached.
AUTO STOP: Automatic stop and reverse motion followed by stop
when the preset torque is reached.
AUTO REVERSE OFF: Auto reverse mode is not activated.
10
CAL
Calibrates the contra-angle to ensure torque accuracy each time
the contra-angle is replaced or lubricated (keep pressed for more
than 2 seconds).
11
MEMO
Saves modifications to the speed, torque limit values and auto
reverse mode in each system where modifications are possible
(keep pressed for more than 2 seconds).
12
SOUND VOLUME
Adjusts the sound volume.
13
AC-IN LAMP
Lights green when the control unit is connected to the mains.
14
CHRG LAMP
Lights or flashes orange while battery is charging or in the refresh
mode (see chapter 6.9. Battery Refresh) and in an error condition
(see chapter 8. Error Code).
Summary of Contents for X-SMART Plus
Page 1: ...USER MANUAL...
Page 3: ...FISDR JAP F1902124 X 00 2011 created 07 2011 4 208...
Page 83: ...FISDR JAP F1902124 X 00 2011 created 07 2011 84 208...
Page 163: ...FISDR JAP F1902124 X 00 2011 created 07 2011 164 208...
Page 207: ...FISDR JAP F1902124 X 00 2011 created 07 2011 208 208...
Page 208: ...FISDR JAP F1902124 X 00 2011 created 07 2011 209 208...