Auto Reverse Off
The motor handpiece stops and does not reverse when the load exceeds the preset
Torque Limit Value. The Torque Limit box will show “S.t”.
If you want the file to rotate forward again, press the
FORWARD
button twice.
If you want the file to rotate in reverse, press and hold the
REVERSE
button twice.
6. Completion of Medical Treatment
• Remove contra-angle attachment for lubrication and sterilization.
• Wipe down motor handpiece with disinfectant as per the cleaning, disinfection, &
sterilization procedure in Section H.
• Finally, return the motor handpiece to the charging stand.
NOTE: If 3 minutes pass without operation of any key the power turns off automatically.
However, the power does not turn off automatically if the motor handpiece is in motion, even if it
is not used.
G. LUBRICATING THE CONTRA-ANGLE ATTACHMENT
•
Lubricate the contra-angle attachment only.
• Lubricate after each use and before sterilization.
• Screw the spray nozzle onto the Spray by approx.10 turns.
• Insert the spray nozzle into the rear part of the contra-angle
attachment and lubricate for 2-3 seconds until oil comes
out from the attachment head.
• Position attachment vertically so all extra oil will drain out,
and then wipe off outside surfaces.
PRECAUTION
• Do not use a spray can upside down. In such case only spray gas comes out instead of oil.
• To prevent attachment flying off from a spray nozzle – hold it securely.
Fig. 12