9. MAINTENANCE 
 

 

 

 

 

 

 

 

/!\ 

ATTENTION:

 

Before carrying out any maintenance operation inside the machine or
any technical activity to the connections, disconnect the electric cable
from the socket; in such a way the machine will be completely insulated
from the electric wiring system. 
If you still have any doubts, difficulties or any possibility of mistake,
contact our Technical Service to avoid any risks or damages. 

 

Cleaning of the working chamber 

It is advisable to keep always clean the internal surface of the working chamber to allow 
the correct refraction of the emitted rays; be careful to use exclusively non-agressive liquid 
detergent specific for lucent metal surfaces. To move freely inside the furnace, we suggest 
to remove the lamps as specified in the following chapter. 

 

Removal or replacement of the ultraviolet lamps 

There are 4 UV lamps totally. In order to detach them, pull softly and unclutch them from 
their sockets (bayonet joint). 
 

Removal or replacement of UV spots  

There are 2 UV spots totally, different by the available spectrum. Both spots are oriented 
so to obtain the best focus in the middle of the rotating plate. The spots are mounted on a 
movable support with flexible clips. To remove the spots, firstly extract the orizontal lamps 
to gain more handling room, then unscrew the nuts which fix the support, and then let 
them gently shift outwards. The wires are plugged to each spot by means of insulated 
terminals. To order spare parts, specify the name of the manufacturer printed on the 
parabola. (Osram or Philips). 

 

 

10. TECHNICAL SPECIFICATIONS 
 

 

Description  

Width 

255 mm 

Depth at the base   

320 mm 

Height 

320 mm 

Net weight (with bell-jar) 

8,3 kg (8,8 kg) 

Gross weight (with bell-jar) 

10,0 kg (10,5 kg) 

Rated voltage 

230 v  AC – 50/60 Hz 

Fuses 

3.15 A 

Absorption 

250 W 

UV lamps 

compact fluorescent lamps  

from 400 to 550 nm 

UV spots 

2 units, 75 W 

(1+1 with different light spectrum) 

from 320 to 550 nm 

Rotating plate 

 80 mm 

 
 
 
 

Summary of Contents for Photopol

Page 1: ...LIGHT CURING UNIT PHOTOPOL Light USER AND MAINTENANCE MANUAL ...

Page 2: ...cessories needed for a correct and practical processing is included in the equipment rotating plate to grant best irradiation of the models sliding table to easily position and extract the objects cooling fan Besides these basic accessories other tools are avilable on request a support for crown and bridges a bell jar for vaccum treatments and a small pump for creating the vaccuum 2 TECHNICAL REFE...

Page 3: ...ay be removed or modified Do not allow unauthorized persons to try to repair the machine 5 INSTALLATION INSTRUCTIONS ATTENTION Installation of this machine is quite easy but it must be carried out paying utmost attention in order to avoid any mistake which may originate problems during operation inconvenience or even damages 1 Place the machine on a suitable work bench in a stable position 2 Check...

Page 4: ...UV spots will switch on the selection is confirmed by the corresponding LED which will lighten by pressing this key all lighting sources will switch on simultaneously UV cold light lamps and UV spots the selection is confirmed by the correspondent LEDs which will lighten by pressing this key you can choose the operating times The data are shown on the display As a standard it is possible to set th...

Page 5: ...ne will warn the user acoustically When the door is opened alla parameters will be set to defaults and the furnace will bre ready for another light curing cycle If needed the cycle can be aborted by holding START for some seconds the display will be reset to zero values and all LEDs will switch off 8 ACCESSORIES AND OPTIONALS Set of cap support A5402K optional consisting of a disc with mobile meta...

Page 6: ...r to detach them pull softly and unclutch them from their sockets bayonet joint Removal or replacement of UV spots There are 2 UV spots totally different by the available spectrum Both spots are oriented so to obtain the best focus in the middle of the rotating plate The spots are mounted on a movable support with flexible clips To remove the spots firstly extract the orizontal lamps to gain more ...

Page 7: ...eholder unit 3 15A IG Main shutter TR Transformer 230v 12v for feeding the spots RE Reactors fo UV cold light lamps L1 4 compact flurescent lamps blue violet color L2 Spot UV blue violet color L3 Spot UV white pink color FAN Cooling fan MS Door microswitch ...

Page 8: ...406013 TABLE SHIFTHING ROD 18 5406012 PAWL WITH DUMMY HOLE 19 5406011 PAWL WITH THROUGH HOLE 20 5406006 REFLECTING SURFACE 21 NEA003 UVA COMPACT BLUE VIOLET LAMP 22 NEC044 SAFETY MICROSWITCH 23 5406008 DOOR ROTATION BUSHING 24 5406025 HIGH PROTECTION ANTI UV SCREEN 25 5406002 DOOR 26 NEC043 CYLINDRICAL MAGNET 27 1051014 PLASTIC RAIL FOR MAGNET 28 5406010L MARK LABEL PHOTOPOL LIGHT 29 5406004 LOWER...

Page 9: ......

Page 10: ...FAN 45 NPR011 STRAIGHT MALE PIPE FITTING 6 1 8 46 NPR206 MALE FEMALE SLEEVE 1 8 47 NEA001 REACTOR FOR COMPACT LAMP 48 NEA005 TRANSFORMER 230V 12V FOR SPOT 49 NEC016 LIGHTED BIPOLAR SWITCH 50 NEA203 TOWER M3 L 30 51 5406100P POWER BOARD PHOTOPOL LIGHT 52 5406100C CONTROL BOARD PHOTOPOL LIGHT 53 NEA202 INSULATING DISTANCE PIECE L 5 ...

Page 11: ......

Page 12: ...DENTALFARM s r l Via Susa 9 a 10138 TORINO ITALY TECHNICAL COMMERCIAL SERVICE 39 011 4346588 AFTER SALE SERVICE 011 4346632 FAX 011 4346366 E mail info dentalfarm it Website www dentalfarm it ...

Reviews: