dentalfarm Photopol Maintenance Manual Download Page 2

1.  DESCRIPTION 

 

PHOTOPOL Light

 has been expressly designed to carry out light curing operations by 

means of ultraviolet rays.   
The system is extremely versatile and can suit the most discerning requirements of both 
the Dental Laboratory and Dental Cabinet. 
 
This machine allows to treat the light-hardening composites: the wide operating spectrum 
assured by this unit (ranging from 320 to 550 nm) permits to treat in the most correct way 
all the products available on the market. 
 
PHOTOPOL inside walls are made of highly reflecting, polished aluminium which ensures 
the ideal distribution of the rays all over the working area. The glass door allows a wide 
opening and it is fitted with a higly-protecting screen permitting to steadily supervise the 
working cycle with no damage for your eyes. 
 
All the functions are controlled and operated by an electronic microprocessor circuit. The 
operating software features a logical user interface providing ease of use and immediate 
understanding. 
 
A set of essential accessories needed for a correct and practical processing is included in 
the equipment: rotating plate to grant best irradiation of the models, sliding table to easily 
position and extract the objects, cooling fan.  
Besides these basic accessories, other tools are avilable on request: a support for crown 
and bridges, a bell-jar for vaccum treatments and a small pump for creating the vaccuum. 
 

 

2. TECHNICAL REFERENCE REGULATIONS AND TEST 
PROCEDURES 

 

The light-curing furnace is mass-manufactured by  DENTALFARM in compliance with 
technical and safety rules in force, as provided for by the2006/42 EEC Community Directive 
on machinery. 
 
Careful inspection and full routine testing is carried out singularly on each machine which is 
furtherly tested by an automatic test installation assuring compliance with the fixed limits. 
 
 

3. DEMOLITION AND WASTE DISPOSAL 

 

According to International regulations, this unit has been classified as AEE 
(electric and electronic device, whose correct operation depends on electric 
currents and electromagnetic fields) and as a consequence, at the end of its 
lifetime, it can not be treated as normal waste material but it must be disposed 
separately, complying with the Directive 2002/96/CE. 
 

 
 
 

 

Summary of Contents for Photopol

Page 1: ...LIGHT CURING UNIT PHOTOPOL Light USER AND MAINTENANCE MANUAL ...

Page 2: ...cessories needed for a correct and practical processing is included in the equipment rotating plate to grant best irradiation of the models sliding table to easily position and extract the objects cooling fan Besides these basic accessories other tools are avilable on request a support for crown and bridges a bell jar for vaccum treatments and a small pump for creating the vaccuum 2 TECHNICAL REFE...

Page 3: ...ay be removed or modified Do not allow unauthorized persons to try to repair the machine 5 INSTALLATION INSTRUCTIONS ATTENTION Installation of this machine is quite easy but it must be carried out paying utmost attention in order to avoid any mistake which may originate problems during operation inconvenience or even damages 1 Place the machine on a suitable work bench in a stable position 2 Check...

Page 4: ...UV spots will switch on the selection is confirmed by the corresponding LED which will lighten by pressing this key all lighting sources will switch on simultaneously UV cold light lamps and UV spots the selection is confirmed by the correspondent LEDs which will lighten by pressing this key you can choose the operating times The data are shown on the display As a standard it is possible to set th...

Page 5: ...ne will warn the user acoustically When the door is opened alla parameters will be set to defaults and the furnace will bre ready for another light curing cycle If needed the cycle can be aborted by holding START for some seconds the display will be reset to zero values and all LEDs will switch off 8 ACCESSORIES AND OPTIONALS Set of cap support A5402K optional consisting of a disc with mobile meta...

Page 6: ...r to detach them pull softly and unclutch them from their sockets bayonet joint Removal or replacement of UV spots There are 2 UV spots totally different by the available spectrum Both spots are oriented so to obtain the best focus in the middle of the rotating plate The spots are mounted on a movable support with flexible clips To remove the spots firstly extract the orizontal lamps to gain more ...

Page 7: ...eholder unit 3 15A IG Main shutter TR Transformer 230v 12v for feeding the spots RE Reactors fo UV cold light lamps L1 4 compact flurescent lamps blue violet color L2 Spot UV blue violet color L3 Spot UV white pink color FAN Cooling fan MS Door microswitch ...

Page 8: ...406013 TABLE SHIFTHING ROD 18 5406012 PAWL WITH DUMMY HOLE 19 5406011 PAWL WITH THROUGH HOLE 20 5406006 REFLECTING SURFACE 21 NEA003 UVA COMPACT BLUE VIOLET LAMP 22 NEC044 SAFETY MICROSWITCH 23 5406008 DOOR ROTATION BUSHING 24 5406025 HIGH PROTECTION ANTI UV SCREEN 25 5406002 DOOR 26 NEC043 CYLINDRICAL MAGNET 27 1051014 PLASTIC RAIL FOR MAGNET 28 5406010L MARK LABEL PHOTOPOL LIGHT 29 5406004 LOWER...

Page 9: ......

Page 10: ...FAN 45 NPR011 STRAIGHT MALE PIPE FITTING 6 1 8 46 NPR206 MALE FEMALE SLEEVE 1 8 47 NEA001 REACTOR FOR COMPACT LAMP 48 NEA005 TRANSFORMER 230V 12V FOR SPOT 49 NEC016 LIGHTED BIPOLAR SWITCH 50 NEA203 TOWER M3 L 30 51 5406100P POWER BOARD PHOTOPOL LIGHT 52 5406100C CONTROL BOARD PHOTOPOL LIGHT 53 NEA202 INSULATING DISTANCE PIECE L 5 ...

Page 11: ......

Page 12: ...DENTALFARM s r l Via Susa 9 a 10138 TORINO ITALY TECHNICAL COMMERCIAL SERVICE 39 011 4346588 AFTER SALE SERVICE 011 4346632 FAX 011 4346366 E mail info dentalfarm it Website www dentalfarm it ...