7.  OPERATIONAL INSTRUCTIONS

 

 

1.  Switch on  the main switch, the display switches on and the unit is ready for the input 

of operating modes and time.  

2.  Open the door and let the working plan shift forward, in order to make the positioning 

of the object on the rotating plate more easy. Then shift it back till the end of its 
stroke, i.e. the best-exposure position. 

3.  Close the door 
4.  Press the function key, choose a function, the LED will be on (the function keys 

exclude themselves reciprocally, so, in case of an incorrect choice, you can correct 
simply by pressing the desired function key.  

5.  Press the keys     to set the time 
6. Press 

START

 
If the door is opened during operation, the cycle will be interrupted and as the door is 
closed, it will re-start from the point it has been stopped.  
If, during the cycle, the cold light lamps have been working, the cycle will stop immediately 
after the set time has elapsed; on the contrary, if during the cycle UV spots have been 
working, there will be a determined extra time for the cooling fan to work, in order to obtain 
a temperature of about 35°C in the working chamber. 
Once the cycle is over, the machine will warn the user acoustically. When the door is 
opened, alla parameters will be set to defaults and the furnace will bre ready for another 
light-curing cycle.  
If needed, the cycle can be aborted by holding 

START

 for some seconds: the display will 

be reset to zero values and all LEDs will switch off. 
 
 

8. ACCESSORIES AND OPTIONALS 

 

Set of cap support

 

A5402K

 (optional) consisting of a disc with mobile metallic pins in 

various sizes, which has been expressly studied to support small pieces. 
 
Vacuum bell-jar code A5406B (optional) allows to perform vacuum treatments, being 
protected from dirty particles. The tool has to be mounted when confirming the order, so 
that it can be integrated with all other functions (rotating plate for uniform exposure, fixed 
distances from light bulbs). 
To use the vaccum bell-jar, place the object on the rotating plate, lock the bell-jar, build up 
vacuum by means of a pump, and then finally start the light-curing cycle. If, after placing 
the bell-jar in place, this does not perfectly adhere to the rubber joint, keep a hand on it for 
a while, gently pressing it manually against the joint.   
  
Vacuum pump MINIVAC A5404 (optional) small in size, this pump features an adequate 
power to create vacuum under the bell-jar. The pump is placed laterally, aside the furnace, 
and is connected to the unit by means of a pipe-fitting. 
 
 

 
 
 

Summary of Contents for Photopol

Page 1: ...LIGHT CURING UNIT PHOTOPOL Light USER AND MAINTENANCE MANUAL ...

Page 2: ...cessories needed for a correct and practical processing is included in the equipment rotating plate to grant best irradiation of the models sliding table to easily position and extract the objects cooling fan Besides these basic accessories other tools are avilable on request a support for crown and bridges a bell jar for vaccum treatments and a small pump for creating the vaccuum 2 TECHNICAL REFE...

Page 3: ...ay be removed or modified Do not allow unauthorized persons to try to repair the machine 5 INSTALLATION INSTRUCTIONS ATTENTION Installation of this machine is quite easy but it must be carried out paying utmost attention in order to avoid any mistake which may originate problems during operation inconvenience or even damages 1 Place the machine on a suitable work bench in a stable position 2 Check...

Page 4: ...UV spots will switch on the selection is confirmed by the corresponding LED which will lighten by pressing this key all lighting sources will switch on simultaneously UV cold light lamps and UV spots the selection is confirmed by the correspondent LEDs which will lighten by pressing this key you can choose the operating times The data are shown on the display As a standard it is possible to set th...

Page 5: ...ne will warn the user acoustically When the door is opened alla parameters will be set to defaults and the furnace will bre ready for another light curing cycle If needed the cycle can be aborted by holding START for some seconds the display will be reset to zero values and all LEDs will switch off 8 ACCESSORIES AND OPTIONALS Set of cap support A5402K optional consisting of a disc with mobile meta...

Page 6: ...r to detach them pull softly and unclutch them from their sockets bayonet joint Removal or replacement of UV spots There are 2 UV spots totally different by the available spectrum Both spots are oriented so to obtain the best focus in the middle of the rotating plate The spots are mounted on a movable support with flexible clips To remove the spots firstly extract the orizontal lamps to gain more ...

Page 7: ...eholder unit 3 15A IG Main shutter TR Transformer 230v 12v for feeding the spots RE Reactors fo UV cold light lamps L1 4 compact flurescent lamps blue violet color L2 Spot UV blue violet color L3 Spot UV white pink color FAN Cooling fan MS Door microswitch ...

Page 8: ...406013 TABLE SHIFTHING ROD 18 5406012 PAWL WITH DUMMY HOLE 19 5406011 PAWL WITH THROUGH HOLE 20 5406006 REFLECTING SURFACE 21 NEA003 UVA COMPACT BLUE VIOLET LAMP 22 NEC044 SAFETY MICROSWITCH 23 5406008 DOOR ROTATION BUSHING 24 5406025 HIGH PROTECTION ANTI UV SCREEN 25 5406002 DOOR 26 NEC043 CYLINDRICAL MAGNET 27 1051014 PLASTIC RAIL FOR MAGNET 28 5406010L MARK LABEL PHOTOPOL LIGHT 29 5406004 LOWER...

Page 9: ......

Page 10: ...FAN 45 NPR011 STRAIGHT MALE PIPE FITTING 6 1 8 46 NPR206 MALE FEMALE SLEEVE 1 8 47 NEA001 REACTOR FOR COMPACT LAMP 48 NEA005 TRANSFORMER 230V 12V FOR SPOT 49 NEC016 LIGHTED BIPOLAR SWITCH 50 NEA203 TOWER M3 L 30 51 5406100P POWER BOARD PHOTOPOL LIGHT 52 5406100C CONTROL BOARD PHOTOPOL LIGHT 53 NEA202 INSULATING DISTANCE PIECE L 5 ...

Page 11: ......

Page 12: ...DENTALFARM s r l Via Susa 9 a 10138 TORINO ITALY TECHNICAL COMMERCIAL SERVICE 39 011 4346588 AFTER SALE SERVICE 011 4346632 FAX 011 4346366 E mail info dentalfarm it Website www dentalfarm it ...

Reviews: