6. DESCRIPTION OF THE WORKING FUNCTIONS
All the functions of the machine can be selected, stored and operated by means of the
electronic keyboard. The key panel has three keys to select operating modes, two keys for
setting the time, and one key for starting the cycle. The function keys operate firm working
programs, which include all necessary operationss, like the rotation of the plate and the
switching on of the working fan.
To start the unit, it is necessary to be aware of the functions operated by each key and the
input sequence.
by pressing this key, the UV cold light lamps will be switched on (the selection
is confirmed by the corresponding LED, which wil lighten):
by pressing this key, the 2 high-intensity UV spots will switch on (the selection
is confirmed by the corresponding LED, which will lighten);
by pressing this key, all lighting sources will switch on simultaneously (UV cold
light lamps and UV spots – the selection is confirmed by the correspondent
LEDs, which will lighten);
by pressing this key, you can choose the operating times.
The data are shown on the display. As a standard, it is
possible to set the minutes (from 1 to 20); should you wish
to set the seconds, as well, push both keys (arrows) and set
the value (from 1 to 59).
Once the operating mode and time have been confirmed, the cycle begins by
pressing START.
/!\
ATTENTION
The more relevant difference betwen the two UV lighting sources is
their intensity: for this reason, please carefully plan the setting of
the operating times. Cycles with cold light lamps can last till 20
minutes, while it is highly recommended not to exceed 6-7 minutes
when operating with spots.