Enable High-Pressure Mode
Section IV Operation
High-pressure mode on the Osprey compressor is
enabled by using the control buttons located on
the C2 smart control panel
(Figure 13)
.
To enable high-pressure mode:
1. Press
NEXT
button to cycle through the menu
screens. Stop at the screen that displays
HI
PRESSURE MODE OFF
(Figure 14)
.
NOTE:
This will require seven to nine presses,
based on the model of Osprey being operated.
2. Press and hold
RESET
button for 10 seconds.
The screen will now display
HI PRESSURE
MODE ON 105 130 PSI
(Figure 15)
.
3. Press
NEXT
button. Screen will display this
warning:
IN HI PRESS MODE OWL COM
DISABLED
. This indicates that the On Wall
Logic (OWL) Smart Hub is in high-pressure
mode and will have limited functionality with
the compressor
(Figure 16)
.
Figure 14. Press NEXT button to cycle to screen
Figure 15. Press and hold RESET button for 10 seconds
Figure 16. Press NEXT button
High-pressure mode should only be activated
when supplying air to a dedicated milling
machine that requires higher pressure.
CAUTION
4. Press
NEXT
button again to return to the main
screen.
Figure 13. Osprey compressor C2 smart control panel
Control
Buttons
Electronic
Display
Summary of Contents for RAM VAC COSP22S
Page 1: ...OSPREY Osprey TM Dental Compressor User Manual ...
Page 3: ...www DentalEZ com 866 DTE INFO ii OspreyTM Compressor PN 1045DOC_E ...
Page 13: ...www DentalEZ com 866 DTE INFO 10 OspreyTM Compressor PN 1045DOC_E ...
Page 27: ...www DentalEZ com 866 DTE INFO 24 OspreyTM Compressor PN 1045DOC_E ...
Page 33: ...www DentalEZ com 866 DTE INFO 30 OspreyTM Compressor PN 1045DOC_E ...
Page 35: ...www DentalEZ com 866 DTE INFO 32 OspreyTM Compressor PN 1045DOC_E ...
Page 37: ...www DentalEZ com 866 DTE INFO 34 OspreyTM Compressor PN 1045DOC_E ...