User Manual
2
EverLight®
(Post Mount)
Section I Introduction
Classifications
Medical - General Medical Equipment
Certified as to electrical shock, fire
and mechanical hazards only in accor
-
dance with UL 60601-1,
CAN/CSA-C22.2 No. 601.1,
CAN/CSA-C22.2 No. 60601-1 (2008) &
ANSI/AAMI ES 60601-1 (2005).
–
Type of protection against electric shock: Class 1
Equipment
–
Degree of protection against the ingress of water:
Ordinary
–
Equipment not suitable for use in the presence of
a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
–
Mode of operation: Continuous
–
Environmental conditions recommended ranges
for:
Transport and Storage:
-
Temperature range of - 10º C to + 70º C
-
Relative humidity range of 10% to 100%
-
Atmospheric pressure range of 500 to 1060 kPa
Normal Use:
-
Temperature range of + 10º C to + 40º C
-
Relative humidity range of 30% to 75%
-
Atmospheric pressure range of 700 to 1060 kPa
Power Supply:
Primary: 100-240V AC, 50/60 Hz.
Secondary: 12V DC
Operating Amperage:
.4 max. - .16 min. Amp
Focal Length:
68.6 cm (27")
Color Temperature:
High: 5800° Kelvin
(Color Matching)
Medium: 4500° Kelvin
(Operative Setting)
Low: 3200° Kelvin
(Composite Setting)
Light Intensity:
High: 2790 Footcandles
(30,000 Lux)
Medium: 2325 Footcandles
(25,000 Lux)
Low: 1395 Footcandles
(15,000 Lux)
Light Pattern:
16.5 cm (6-1/2") x 8.9 cm (3-1/2")
Unit Weight:
Post Mount:
12 kg (27 lbs.)
The authorized European representative is:
DentalEZ (GB) Ltd.
Cleveland Way
Hemel Hempstead, Hertfordshire
HP2 7DY, England
Phone: (01442) 269301
Attn: Mr. Andy King
This manual contains installation, operation and care
instructions and user service information for the
EverLight
®
LED operatory light.
The EverLight operatory light system is intended to
be used by trained professional dental care personnel
to provide shadow free illumination to the oral cavity
for dental practices. The operatory light arm system
allows optimal light positioning, which for a reclined
patient, is normally from the sagittal plane directly
above or slightly behind the patient's oral cavity. The
operator will typically be positioned behind or to one
side of the patient. The arm assembly is counterbal
-
anced so that minimum force is required to move
the light head assembly. The operatory light is not
designed to handle any external loads, such as hanging
foreign objects from the light system.
The EverLight operatory light is designed to provide
trouble-free service when installed, operated and
maintained according to the procedures set forth in
this manual.
To ensure correct installation, carefully read all
the instructions contained in this manual
paying close attention to all notes, notices,
cautions and warnings.
After the light is installed, please review the operation
procedures and care guidelines with the doctor's staff.
Then
leave this manual in the doctor's office
for
future reference.
53HN
Specifications