17
EMC Information
Table 4
Recommended separation distance between
Portable and mobile RF communications equipment and the model @ 3Vrms
The
EverLight
is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the
EverLight
can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the
EverLight
as recommended below, according
to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter m
Rated maximum output
power of transmitter W
150 kHz to 80 MHz
P
v
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5
.
3
80 MHz to 800 MHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
5
.
3
800 MHz to 2.5 GHz
P
E
d
⎥
⎦
⎤
⎢
⎣
⎡
=
1
7
0.01 0.12
0.12
0.23
0.1 0.34
0.34
0.74
1 1.7
1.7
2.3
10 3.7
3.7
7.4
100 11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 1:These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.